Wavelength selection for low-saturation pulse oximetry
- 1 March 1997
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 44 (3) , 148-158
- https://doi.org/10.1109/10.554761
Abstract
Conventional pulse oximeters are accurate at high oxygen saturation under a variety of physiological conditions but show worsening accuracy at lower saturation (below 70%). Numerical modeling suggests that sensors fabricated with 735 and 890 nm emitters should read more accurately at low saturation under a variety of conditions than sensors made with conventionally used 660 and 900 nm band emitters. Recent animal testing confirms this expectation. It is postulated that the most repeatable and stable accuracy of the pulse oximeter occurs when the fractional change in photon path lengths due to perturbations in the tissue (relative to the conditions present during system calibration) is equivalent at the two wavelengths. Additionally, the penetration depth (and/or breadth) of the probing light needs to be well matched at the two wavelengths in order to minimize the effects of tissue heterogeneity. At high saturation these conditions are optimally met with 660 and 900 nm band emitters, while at low saturation 735 and 890 nm provide better performance.Keywords
This publication has 19 references indexed in Scilit:
- Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetryIEEE Transactions on Biomedical Engineering, 1991
- Effect of anemia on pulse oximeter accuracy at low saturationJournal of Clinical Monitoring and Computing, 1990
- A NON-INVASIVE REFLECTANCE OXIMETER AS A USEFUL MONITOR IN ICUAnesthesiology, 1989
- Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical propertiesApplied Optics, 1989
- Errors in 14 pulse oximeters during profound hypoxiaJournal of Clinical Monitoring and Computing, 1989
- Statistics of Penetration Depth of Photons Re-emitted from Irradiated TissueJournal of Modern Optics, 1989
- Accuracy of Pulse Oximetry During Arterial Oxyhemoglobin Desaturation in DogsAnesthesiology, 1988
- Model for photon migration in turbid biological mediaJournal of the Optical Society of America A, 1987
- Percentiles for gas values in human umbilical cord bloodEuropean Journal of Obstetrics & Gynecology and Reproductive Biology, 1983
- Real Time OximetryPublished by Springer Nature ,1983