A Monte Carlo calculation of dosimetric parameters of 90Sr/90Y and 192Ir SS sources for intravascular brachytherapy
- 9 November 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (11) , 2528-2535
- https://doi.org/10.1118/1.1319374
Abstract
The AAPM TG-60 report has proposed various dose calculation parameters for intravascular brachytherapy (IVBT). These parameters include the dose rate constant (or the dose rate at a reference position for a beta-particle emitting source), the radial dose function, and the anisotropy function. In this work, we have used a modified EGS4 Monte Carlo system to calculate these parameters for the two most commonly used IVBT sources (the beta-particle emitting 90Sr/90Y source and the photon emitting 192Ir SS source). To ensure the calculation accuracy, the present calculation was compared with several measurements and calculations reported by other authors. Excellent agreement was found for the results with the photon source. For beta-particle source calculation, the present results for a variety of point sources agree very well with a previous work. The presently calculated radial dose functions for the 90Sr/90Y source are consistent with those of a published work for intermediate radial distances. The dose uniformity in the axial direction was also studied. The contributions of bremsstrahlung photons to total doses for the 90Sr/90Y beta source, and the influence of ignoring electron transport on calculated doses for the 192Ir SS photon source are discussed.Keywords
This publication has 15 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60Medical Physics, 1999
- Combined experimental and Monte Carlo verification of brachytherapy plans for vaginal applicatorsPhysics in Medicine & Biology, 1998
- Monte Carlo dosimetry of the VariSource high dose rate 192Ir sourceMedical Physics, 1998
- Calibration and characterization of beta-particle sources for intravascular brachytherapyMedical Physics, 1998
- EGS4 dosimetry calculations for cylindrically symmetric brachytherapy sourcesMedical Physics, 1996
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Comparison of measured and calculated dose rates in water near I‐125 and Ir‐192 seedsMedical Physics, 1991
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986
- A short atlas of beta-ray spectraPhysics in Medicine & Biology, 1983