Cooled Intramural Needle Catheter Ablation Creates Deeper Lesions than Irrigated Tip Catheter Ablation
- 22 July 2004
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 27 (7) , 965-970
- https://doi.org/10.1111/j.1540-8159.2004.00566.x
Abstract
Endocardial radiofrequency ablation of the left ventricle does not create transmural lesions reliably even with active electrode cooling. The authors developed a prototype catheter with an internally cooled needle electrode that could be advanced an adjustable distance into the myocardium. Freshly excised hearts from eight male sheep were perfused and superfused using oxygenated ovine blood. Ablations were performed for 2 minutes using the prototype catheter and a conventional endocardial 5-mm irrigated tip ablation catheter at target temperatures of 80 degrees C and 50 degrees C, respectively. The prototype catheter needle was inserted 12 mm deep for all ablations. The maximal power and irrigation rate was 50 W, 20 mL/min for the irrigated tip catheter and 20 W, 10 mL/min for the intramural needle catheter. Intramural needle lesions were significantly deeper (13.5 +/- 2.3 vs 9.1 +/- 1.3 mm, P < 0.01) but less wide (8.7 +/- 1.5 vs 12.7 +/- 1.9 mm, P < 0.01) than irrigated tip lesions. Popping occurred during 12 (37%) of the 32 irrigated tip ablations. Popping did not occur during intramural needle ablation. The cooled intramural needle ablation catheter creates lesions that are significantly deeper than irrigated tip catheters with less tissue boiling. In contrast to irrigated tip ablation, electrode temperature monitoring can be used to determine if a lesion has been created during intramural needle ablation. The cooled intramural needle ablation lesions were of a clinically useful width, addressing one of the main recognized deficiencies of intramural needle ablation.Keywords
This publication has 19 references indexed in Scilit:
- Prospective Randomized Comparison of Irrigated-Tip Versus Conventional-Tip Catheters for Ablation of Common FlutterCirculation, 2000
- A new catheter design using needle electrode for subendocardial RF ablation of ventricular muscles: finite element analysis and in vitro experimentsIEEE Transactions on Biomedical Engineering, 2000
- Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine MyocardiumCirculation, 1999
- A cooled needle electrode for radiofrequency tissue ablation: Thermodynamic aspects of improved performance compared with conventional needle designAcademic Radiology, 1996
- Radiofrequency Delivery Through a Cooled Catheter Tip Allows the Creation of Larger Endomyocardial Lesions in the Ovine HeartJournal of Cardiovascular Electrophysiology, 1995
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Development of a new intraoperative radiofrequency ablation technique using a needle electrodeThe Annals of Thoracic Surgery, 1994
- The Biophysics of Radiofrequency Catheter Ablation in the Heart: The Importance of Temperature MonitoringPacing and Clinical Electrophysiology, 1993
- Functional role of the epicardium in postinfarction ventricular tachycardia. Observations derived from computerized epicardial activation mapping, entrainment, and epicardial laser photoablation.Circulation, 1991
- The Effect of Electrode Design on the Efficiency of Delivery of Radiofrequency Energy to Cardiac Tissue In VitroPacing and Clinical Electrophysiology, 1989