Monitoring cardiac output using the femoral and radial arterial pressure waveform
- 26 June 2006
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 61 (8) , 743-746
- https://doi.org/10.1111/j.1365-2044.2006.04712.x
Abstract
This study was performed to determine the interchangeability of femoral artery pressure and radial artery pressure measurements as the input for the PiCCO system (Pulsion Medical Systems, Munich, Germany). We studied 15 intensive care patients following cardiac surgery. Five-second averages of the cardiac output derived from the femoral artery pressure (COfem) were compared to 5-s averages derived from the radial artery pressure (COrad). One patient was excluded due to problems in the pattern recognition of the arterial pressure signal. In the remaining 14 patients, 14 734 comparative cardiac output values were analysed. The mean sample time was 88 min, range [30-119 min]. Mean (SD) COfem was 6.24 (1.1) l.min(-1) and mean COrad 6.23 (1.1) l.min(-1). Bland-Altman analysis showed an excellent agreement with a bias of - 0.01 l.min(-1), and limits of agreement from 0.60 to - 0.62 l.min(-1). If changes in CO were > 0.5 l.min(-1), the direction of changes in COfem and COrad were equal in 97% of instances. We conclude that femoral artery pressure and radial artery pressure are interchangeable as inputs for the PiCCO device.Keywords
This publication has 15 references indexed in Scilit:
- Measurement of cardiac output by transpulmonary arterial thermodilution using a long radial artery catheter. A comparison with intermittent pulmonary artery thermodilutionAnaesthesia, 2004
- Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2003
- Propofol or sevoflurane anesthesia without muscle relaxants allow the early extubation of myasthenic patientsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2003
- Can a clinically useful aortic pressure wave be derived from a radial pressure wave?British Journal of Anaesthesia, 2002
- Functional hemodynamic monitoringIntensive Care Medicine, 2002
- Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb,British Journal of Anaesthesia, 2001
- Continuous, Less Invasive, Hemodynamic Monitoring in Intensive Care After Cardiac SurgeryThe Thoracic and Cardiovascular Surgeon, 1998
- Continuous cardiac output monitoring by pulse contour during cardiac surgeryEuropean Heart Journal, 1990
- Radial Artery-to-Aorta Pressure Difference after Discontinuation of Cardiopulmonary BypassAnesthesiology, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986