Central venous catheters—the inability of ‘intra-atrial ECG’ to prove adequate positioning
Open Access
- 25 June 2004
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 93 (2) , 193-198
- https://doi.org/10.1093/bja/aeh191
Abstract
Background. The classic increase in P wave size, known as ‘P-atriale’, is a widely accepted criterion for determination of proper positioning of central venous catheter tips. Recent transoesophageal echocardiography (TOE) studies did not confirm intra-atrial position despite advancing the central venous catheter further than indicated by ECG guidance. We postulate that the pericardial reflection rather than the entry into the right atrium corresponds to the ECG changes. In order to test our hypothesis we sought to determine the anatomical substrate for the electrical changes in an animal study. Subsequently, a modified version of the study was undertaken in man and is also reported. Methods. In six juvenile pigs the left external jugular vein and right carotid artery were cannulated. A triple-lumen central venous catheter was positioned by ECG guidance using a Seldinger wire as an exploring electrode. The venous and arterial catheters were suture fixed 2 cm beyond the onset of an increase in P wave size. The corresponding anatomical catheter tip position was determined by open exploration of the vessels and the heart. Subsequently the catheter tip position (during advancement) of a pulmonary artery catheter and the corresponding electrical ECG changes were examined in 10 patients during open chest cardiac surgery. Results. All catheters—arterial and venous, in animals and humans—revealed an increase in size of the P wave as well as the QRS complex. All venous catheters were positioned in the superior vena cava, beyond the pericardial reflection but outside the right atrium. All arterial catheters were positioned in the ascending aorta thus also beyond the pericardial reflection. Conclusions. The start of an increase in P wave size does not correspond with the entrance of the right atrium. The anatomic equivalent for the electrophysiological changes of the ECG is the pericardial reflection. ECG guidance is unable to distinguish between venous and arterial catheter position.Keywords
This publication has 13 references indexed in Scilit:
- Intra-atrial ECG is not a reliable method for positioning left internal jugular vein cathetersBritish Journal of Anaesthesia, 2003
- Der Cava-Katheter aus medico-legaler SichtAINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie, 1998
- Central venous catheter placement using the ECG-guided cavafix-certodyn SD catheterJournal of Clinical Anesthesia, 1994
- Accurate placement of central venous cathetersCritical Care Medicine, 1993
- Verification of the position of a central venous catheter by intra‐atrial ECG. When does this method fail?Acta Anaesthesiologica Scandinavica, 1993
- CVP catheter electrocardiography: an alternative to radiographic control after cannulation of central veins?Acta Anaesthesiologica Scandinavica, 1991
- In vitro evaluation of relative perforating potential of central venous catheters: Comparison of materials, selected models, number of lumens, and angles of incidence to simulated membraneJournal of Clinical Monitoring and Computing, 1991
- Complications Associated with Central Venous CathetersChest, 1988
- Impending catheter perforation of superior vena cava: radiographic recognitionAmerican Journal of Roentgenology, 1986
- How Much do Arm Movements Displace Cubital Central Venous Catheters?Acta Anaesthesiologica Scandinavica, 1982