Use of scattered radiation for absolute x-ray energy calibration
- 1 January 1999
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 38-40
- https://doi.org/10.1063/1.1149539
Abstract
A small, simple, and effective x-ray energy monitor has been built for use in energy-sensitive experiments where normal transmission monitoring is not an option. The instrument uses x raysscattered elastically from air, thin polymer films, or other featureless scatterers as a secondary source for measuring the transmission versus energy through a calibrated x-ray spectroscopy standard. In general, this type of energy monitor can be made from many combinations of detectors and scatterers, including ultrahigh vacuum compatible components. We show experimental results from a miniature, p-i-n diode-based monitor, which fits unobtrusively anywhere along the optical path of the experiment, and can be made easily from inexpensive and readily available parts.Keywords