Method of identifying dynamic multileaf collimator irradiation that is highly sensitive to a systematic MLC calibration error
- 9 November 2001
- journal article
- Published by Wiley in Medical Physics
- Vol. 28 (11) , 2220-2226
- https://doi.org/10.1118/1.1408284
Abstract
In intensity modulated radiotherapy (IMRT), radiation is delivered in a multiple of multileaf collimator (MLC) subfields. A subfield with a small leaf‐to‐leaf opening is highly sensitive to a leaf‐ positional error. We introduce a method of identifying and rejecting IMRT plans that are highly sensitive to a systematic MLC gap error (sensitivity to possible random leaf‐positional errors is not addressed here). There are two sources of a systematic MLC gap error: centerline mechanical offset (CMO) and, in the case of a rounded end MLC, radiation field offset (RFO). In IMRT planning system, using an incorrect value of RFO introduces a systematic error ΔRFO that results in all leaf‐to‐leaf gaps that are either too large or too small by (2⋅ΔRFO), whereas assuming that CMO is zero introduces systematic error ΔCMO that results in all gaps that are too large by ΔCMO=CMO. We introduce a concept of the average leaf pair Opening (ALPO) that can be calculated from a dynamic MLC delivery file. We derive an analytic formula for a fractional average fluence error resulting from a systematic gap error of and show that it is inversely proportional to ALPO; explicitly it is equal to in which is generally of the order of 1 mm and This analytic relationship is verified with independent numerical calculations.Keywords
This publication has 16 references indexed in Scilit:
- A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assuranceMedical Physics, 2000
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- An optimized leaf-setting algorithm for beam intensity modulation using dynamic multileaf collimatorsPhysics in Medicine & Biology, 1998
- Testing of dynamic multileaf collimationMedical Physics, 1996
- Clinical implementation of a commercial multileaf collimator: Dosimetry, networking, simulation, and quality assuranceInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimationRadiotherapy and Oncology, 1994
- The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimationPhysics in Medicine & Biology, 1992
- Optimization of stationary and moving beam radiation therapy techniquesRadiotherapy and Oncology, 1988