Determination and correction of quadrature fringe measurement errors in interferometers
- 1 October 1981
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 20 (19) , 3382-3384
- https://doi.org/10.1364/ao.20.003382
Abstract
The precision and accuracy of interferometers using quadrature fringe detection are often limited not by the interferometer itself but by the detector system. There are three typical errors: unequal gain in the two channels; quadrature phase shift error; and zero offsets. This paper describes a simple method for determining the quadrature errors from experimental data obtained in the interferometer and correcting for them. A numerical example demonstrating the significant improvement in the precision of interferometer data is given.Keywords
This publication has 1 reference indexed in Scilit:
- Ultrasonic manometers for low and medium vacua under development at the National Bureau of StandardsJournal of Vacuum Science and Technology, 1977