Comparative Study of Thyroxine Assays Employing Kit Radio-Ligand Reagents
- 1 January 1973
- journal article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 10 (1-6) , 14-22
- https://doi.org/10.1177/000456327301000103
Abstract
A comparative study of some ‘kit’ radiometric procedures for the determination of thyroxine in serum is reported. Four manufacturers' competitive-binding reagents and methods—three incorporating prior extraction and one employing column elution—are examined. Considerable discrepancies exist between the manufacturers' ‘standards’ which, in the main, account for the differences in results obtained by the different methods. Other deficiencies include susceptibility to metal ion interference, variation (up to 8%) in thyroxine extraction and non-specific, i.e. ‘mimic T-4’, measurement. Radio-ligand reagent ‘kits’ need to be critically assessed before being used routinely. Separation of hypo-and hyperthyroid from euthyroid patients is achieved as well with some competitive protein-binding T-4 methods as with serum PBI.Keywords
This publication has 7 references indexed in Scilit:
- Evaluation of a Simplified Technique for the Specific Measurement of Serum Thyroxine ConcentrationJournal of Clinical Endocrinology & Metabolism, 1971
- EVALUATION OF A SIMPLIFIED METHOD FOR DETERMINING SERUM THYROXINE BY COMPETITIVE PROTEIN BINDING ANALYSISActa Endocrinologica, 1970
- Determination of serum thyroxine using a resin sponge techniqueJournal of Clinical Pathology, 1967
- Albumin and “Total Globulin” Fractions of BloodPublished by Elsevier ,1966
- A Study of the Accuracy and Precision of Clinical Chemistry Determinations in 170 Canadian LaboratoriesClinical Chemistry, 1963
- A study of the alkaline ashing method for determination of protein-bound iodine in serumClinica Chimica Acta; International Journal of Clinical Chemistry, 1960
- THE CLINICAL DETERMINATION OF PROTEIN-BOUND IODINE 1Journal of Clinical Investigation, 1951