Comparison of a highly automated 5-h susceptibility testing system, the Cobas-Bact, with two reference methods: Kirby-Bauer disk diffusion and broth microdilution
- 1 December 1987
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 25 (12) , 2372-2377
- https://doi.org/10.1128/jcm.25.12.2372-2377.1987
Abstract
The results of susceptibility tests performed with the Cobas-Bact system were compared with those of the Kirby-Bauer disk diffusion and the broth microdilution methods. The evaluation included tests with 24 antibiotics against 250 isolates of the family Enterobacteriaceae and 13 antibiotics against 100 gram-positive cocci. Complete agreements between the Cobas-Bact and Kirby-Bauer methods were 82.8 and 84.5% for gram-positive cocci and gram-negative bacilli, respectively. Agreements between the Cobas-Bact and broth microdilution methods were 76.7% for gram-positive cocci and 84.8% for gram-negative bacilli. Complete agreements between the Kirby-Bauer and broth microdilution methods were 87.0% for gram-positive cocci and 92.2% for gram-negative bacilli. Despite generally satisfactory results with most organism-antibiotic combinations tested, additional modifications of the Cobas-Bact system are required to reduce the number of major and very major discrepancies, as well as to permit testing of Pseudomonas spp. and other gram-negative nonfermentative bacilli.This publication has 25 references indexed in Scilit:
- Evaluation of the Cobasbact system for rapid antimicrobial susceptibility testing of positive blood culture brothsEuropean Journal of Clinical Microbiology & Infectious Diseases, 1985
- Comparison of the Cathra Repliscan II, the AutoMicrobic system Gram-Negative General Susceptibility-Plus Card, and the Micro-Media System Fox Panel for dilution susceptibility testing of gram-negative bacilliJournal of Clinical Microbiology, 1985
- Evaluation of the Cobasbact automated antimicrobial susceptibility testing systemEuropean Journal of Clinical Microbiology & Infectious Diseases, 1985
- Comparison of two automated instrument systems for rapid susceptibility testing of gram-negative bacilliJournal of Clinical Microbiology, 1983
- Comparison of three automated systems for antimicrobial susceptibility testing of gram-negative bacilliJournal of Clinical Microbiology, 1982
- Evaluation of the Sceptor microdilution antibiotic susceptibility testing system: a collaborative investigationJournal of Clinical Microbiology, 1981
- Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative studyJournal of Clinical Microbiology, 1980
- Evaluation of the sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative studyJournal of Clinical Microbiology, 1980
- Evaluation of the Micro-Media System for Quantitative Antimicrobial Drug Susceptibility Testing: a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1978
- Laboratory Evaluation of a Rapid, Automated Susceptibility Testing System: Report of a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1975