MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration
Top Cited Papers
Open Access
- 16 October 2008
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 49 (11) , 1875-1883
- https://doi.org/10.2967/jnumed.107.049353
Abstract
For quantitative PET information, correction of tissue photon attenuation is mandatory. Generally in conventional PET, the attenuation map is obtained from a transmission scan, which uses a rotating radionuclide source, or from the CT scan in a combined PET/CT scanner. In the case of PET/MRI scanners currently under development, insufficient space for the rotating source exists; the attenuation map can be calculated from the MR image instead. This task is challenging because MR intensities correlate with proton densities and tissue-relaxation properties, rather than with attenuation-related mass density. Methods: We used a combination of local pattern recognition and atlas registration, which captures global variation of anatomy, to predict pseudo-CT images from a given MR image. These pseudo-CT images were then used for attenuation correction, as the process would be performed in a PET/CT scanner. Results: For human brain scans, we show on a database of 17 MR/CT image pairs that our method reliably enables estimation of a pseudo-CT image from the MR image alone. On additional datasets of MRI/PET/CT triplets of human brain scans, we compare MRI-based attenuation correction with CT-based correction. Our approach enables PET quantification with a mean error of 3.2% for predefined regions of interest, which we found to be clinically not significant. However, our method is not specific to brain imaging, and we show promising initial results on 1 whole-body animal dataset. Conclusion: This method allows reliable MRI-based attenuation correction for human brain scans. Further work is necessary to validate the method for whole-body imaging.Keywords
This publication has 25 references indexed in Scilit:
- Simultaneous PET-MRI: a new approach for functional and morphological imagingNature Medicine, 2008
- MR-based attenuation correction for torso-PET/MR imaging: pitfalls in mapping MR to CT dataEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Is MR-guided Attenuation Correction a Viable Option for Dual-Modality PET/MR Imaging?Radiology, 2007
- PET/MR Images Acquired with a Compact MR-compatible PET Detector in a 7-T MagnetRadiology, 2007
- Method for transforming CT images for attenuation correction in PET/CT imagingMedical Physics, 2006
- MRI-Guided SPECT Perfusion Measures and Volumetric MRI in Prodromal Alzheimer DiseaseArchives of Neurology, 2003
- Absolute activity quantitation from projections using an analytical approach: comparison with iterative methods in Tc-99m and I-123 brain SPECTIEEE Transactions on Nuclear Science, 2001
- Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithmIEEE Transactions on Medical Imaging, 2001
- Watershed-based segmentation of 3D MR data for volume quantizationMagnetic Resonance Imaging, 1996
- Segmented MR images for brain attenuation correction in PETPublished by SPIE-Intl Soc Optical Eng ,1994