Recommendations for Intracranial Pressure Monitoring Technology
- 1 June 2000
- journal article
- guideline
- Published by Mary Ann Liebert Inc in Journal of Neurotrauma
- Vol. 17 (6_7) , 497-506
- https://doi.org/10.1089/neu.2000.17.497
Abstract
In patients who require ICP monitoring, a ventricular catheter connected to an external strain gauge transducer or catheter tip pressure transducer device is the most accurate reliable method of monitoring ICP and enables therapeutic CSF drainage. Clinically significant infections or hemorrhage associated with ICP devices causing patient morbidity are rare and should not deter the decision to monitor ICP. Parenchymal catheter tip pressure transducer devices measure ICP similar to ventricular ICP pressure but have the potential for significant measurement differences and drift due to the inability to recalibrate. These devices are advantageous when ventricular ICP is not obtained or if there is obstruction in the fluid couple. Subarachnoid or subdural fluid coupled devices and epidural ICP devices are currently less accurate.Keywords
This publication has 38 references indexed in Scilit:
- Reliability of Intraventricular Pressure Measurement with Fiberoptic or Solid-state Transducers: Avoidance of a Methodological ErrorNeurosurgery, 1998
- A simple technique to limit ICP catheter infectionBritish Journal Of Neurosurgery, 1997
- Pressure-controlled Drainage of Cerebrospinal Fluid: Clinical Experience with a New Type of Ventricular Catheter (Ventcontrol MTC) and an Integrated Piezo-resistive Sensor At Its Tip: Technical NoteNeurosurgery, 1996
- A prospective comparison of fibre-optic and fluid-filled single lumen bolt subdural pressure transducers in ventilated neurosurgical patientsBritish Journal Of Neurosurgery, 1996
- A comparison of extradural and intraparenchymatous intracranial pressures in head injured patientsIntensive Care Medicine, 1995
- Accurate placement of coronal ventricular catheter using stereotactic coordinate-guided free-hand passageJournal of Neurosurgery, 1994
- Relative risks of ventriculostomy infection and morbidityActa Neurochirurgica, 1994
- Intracranial pressure monitoring: Fiberoptic monitor compared with the ventricular catheterSurgical Neurology, 1992
- Continuous monitoring of cerebral tissue pressure in neurosurgical practice—experiences with 100 patientsIntensive Care Medicine, 1990