Volume Ventilation of Infants with Congenital Heart Disease: A Comparison of Dräger, NAD 6000 and Siemens, Servo 900C Ventilators
- 1 January 2001
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 92 (1) , 76-79
- https://doi.org/10.1097/00000539-200101000-00015
Abstract
We compared the ventilation and pulmonary mechanics produced by a new anesthesia ventilator (NAD 6000) using a circle system with that produced by a critical care ventilator (Servo 900C) using a nonrebreathing circuit in infants with congenital heart disease. Twenty patients, aged 1 day to 7 mo, weighing 2.1 to 4.6 kg, were studied. The NAD 6000 had improved alveolar ventilation: Paco2 43 ± 8 vs 47 ± 5 mm Hg (P = 0.005), end-tidal CO2 34 ± 7 vs 37 ± 5 mm Hg (P = 0.042); larger inspired tidal volumes 12.9 ± 2.8 vs 11.3 ± 2.2 mL/kg (P < 0.001), but with higher mean airway pressures 9.7 ± 1.6 vs 8.6 ± 1.3 cm H2O (P < 0.001). These differences in ventilation and airway pressures were not clinically significant. Although there were differences in observed ventilatory variables, both machines provided adequate ventilation when set in the volume control mode. We compared two ventilators for use in infants. Twenty infants undergoing surgery for congenital heart defects were randomized to receive ventilation first with one ventilator, then with the other. Although there were differences in observed ventilatory variables, both machines provided adequate ventilation when set in the volume control mode.Keywords
This publication has 10 references indexed in Scilit:
- Comparison of NAD 6000 and Servo 900C Ventilators in an Infant Lung ModelAnesthesia & Analgesia, 2000
- A Comparison of Three Modes of Ventilation with the Use of an Adult Circle System in an Infant Lung ModelAnesthesia & Analgesia, 1998
- The effect of circuit compliance on delivered ventilation with use of an adult circle system for time cycled volume controlled ventilation using an infant lung modelPediatric Anesthesia, 1998
- Accuracy of Volume Measurements in Mechanically Ventilated Newborns: A Comparative Study of Commercial DevicesJournal of Clinical Monitoring and Computing, 1998
- Increased respiratory system resistance and bronchial smooth muscle hypertrophy in children with acute postoperative pulmonary hypertension.American Journal of Respiratory and Critical Care Medicine, 1995
- In vitro assessment of infant pulmonary function equipmentPediatric Pulmonology, 1995
- Pressure and Flow Limitations of Anesthesia VentilatorsAnesthesiology, 1989
- Perioperative Pulmonary Function in Acute Respiratory FailureAnesthesiology, 1989
- Lung mechanics in congenital heart disease with increased and decreased pulmonary blood flowThe Journal of Pediatrics, 1977
- Lung Mechanics in Normal Infants and Infants with Congenital Heart DiseaseArchives of Disease in Childhood, 1972