Collaborative study to recalibrate the International Reference Preparation of Anti-D Immunoglobulin.
Open Access
- 1 March 1980
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 33 (3) , 249-253
- https://doi.org/10.1136/jcp.33.3.249
Abstract
A collaborative study involving nine independent assays by eight laboratories has recalibrated the anti-D concentration of the International Reference Preparation of Anti-D Immunoglobulin (68/417) in terms of the International Standard for Anti-Rho (anti-D) Incomplete Blood Typing Serum (64/16). This study was carried out when it was found that 68/419 had been calibrated not against 64/16, as originally intended, but inadvertently against another preparation. Based on the results, a revised rounded off value of 300 IU anti-D per ampoule of 68/419 was assigned by the Expert Committee on Biological Standardisation of WHO at its 30th meeting. Many of the known variables in anti-D quantitation using the AutoAnalyzer were considered in the preparation of the protocol for this study. The remarkably close agreement of the results indicated that the format can be used as an acceptable model for interlaboratory studies in the future.This publication has 11 references indexed in Scilit:
- International Collaborative Study of Assay of Anti‐D (Anti‐Rho) ImmunoglobulinBritish Journal of Haematology, 1978
- An Inhibitor to Erythrocyte Agglutination in Bovine Albumin PreparationsVox Sanguinis, 1975
- The effect of ambient temperature on the anti-D assay using the Auto AnalyzerJournal of Clinical Pathology, 1974
- Observations on the Reproducibility of the Bromelised Test Cell Anti‐D Assay Using the Auto‐Analyser1Vox Sanguinis, 1974
- Manipulative and inherent errors in anti-D quantitation using the AutoAnalyzerJournal of Clinical Pathology, 1972
- Reappraisal of an Auto Analyzer Haemagglutination System for Anti‐D QuantificationScandinavian Journal of Haematology, 1972
- An Evaluation of the Single-Channel Auto Analyzer in the Quantitation of Anti-Rh0Vox Sanguinis, 1970
- Automated detection of blood group antibodies.1968
- The Anti-D Content of IgG Preparations for Use in the Prevention of Rh Haemolytic DiseaseVox Sanguinis, 1968
- The International Standard for anti-Rh (anti-D) Incomplete Blood-Typing Serum.1967