Sequential Pulse Countershock Between Two Transvenous Catheters: Feasibility, Safety, and Efficacy
- 1 December 1989
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 12 (12) , 1869-1877
- https://doi.org/10.1111/j.1540-8159.1989.tb01878.x
Abstract
YEE, R., et al.: Sequential Pulse Countershock Between Two Transvenous Catheters: Feasihility, Safety, and Efficacy We evalualed the feasibility, safety, and efficacy of sequential pulse countershock (SqCS) delivered solely through two endocardial catheters for the termination of ventricular tachycardia (VT) and fibrillation (VF) in patients undergoing electrophysiology studies (EPS). Thirty‐four patients (31 men, 3 women) with a mean age of 56.8 ± 10.1 years were studied. Etiology of VT/VF was ischemic heart disease (n = 26), cardiomyopathy (4) repaired tetralogy of Fallot (n = 1), heart transplant (n = 1), and no identifiable heart disease (n = 2). Catheters were positioned successfully in 29 patients. These were positioned in the right ventricular apex (RVA) and the coronary sinus (CS), respectively. The RVA electrode served as the common cathode for both pulses. The two electrodes located near the right atrium/superior vena cava junction served as anode for pulse 1 while the distal CS electrodes served as anode for pulse 2. Twenty‐nine induced VT episodes with cycle length (CL) 220–370 msec were treated. SqCS successfully terminated 15 VT (100–500V) while 14 were accelerated or degenerated to VF. VTCL was longer in success ful SqCS episodes than in those that were accelerated (285 ± 17.3 vs 245 ± 30.8 msec, P < .003). Of 26 VF episodes, 21 were terminated with SqCS (500–900V) and 5 were terminated by transthoracic rescue shocks. On 2 occasions, failure to defibrillate was attributable to poor catheter position at the time of shock. No complications occurred. We conclude that SqCS delivered solely between endocardial catheter electrodes is feasible and effective using energy doses within the range of existing implantable cardioverter defibrillators.Keywords
This publication has 15 references indexed in Scilit:
- Sequential pulse defibrillation in humans: Orthogonal sequential pulse defibrillation with epicardial electrodesJournal of the American College of Cardiology, 1988
- Internal Ventricular Defibrillation with Sequential Pulse Countershock in Pigs: Comparison with Single Pulses and Effects of Pulse SeparationPacing and Clinical Electrophysiology, 1987
- Double and triple sequential shocks reduce ventricular defibrillation threshold in dogs with and without myocardial infarctionJournal of the American College of Cardiology, 1986
- The implanted defibrillator: relation of defibrillating lead configuration and clinical variables to defibrillation thresholdJournal of the American College of Cardiology, 1985
- Improved internal defibrillation with twin pulse sequential energy delivery to different lead orientations in pigsThe American Journal of Cardiology, 1985
- Low energy counterchock using an intravascular catheter in an acute cardiac care settingThe American Journal of Cardiology, 1982
- Clinical transvenous cardioversion of recurrent life-threatening ventricular tachyarrhythmias: Low energy synchronized cardioversion of ventricular tachycardia and termination of ventricular fibrillation in patients using a catheter electrodeAmerican Heart Journal, 1982
- Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human BeingsNew England Journal of Medicine, 1980
- Ventricular defibrillation in the dog with a bielectrode intravascular catheterArchives of internal medicine (1960), 1973
- TRANSVENOUS AUTOMATIC DEFIBRILLATOR — PRELIMINARY CLINICAL TESTS OF ITS DEFIBRILLATING SUBSYSTEMAsaio Journal, 1972