Therapy imaging: Source sizes of radiotherapy beams
- 1 July 1988
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 15 (4) , 517-524
- https://doi.org/10.1118/1.596295
Abstract
We have developed a novel method, which employs large lead collimators and computed tomography reconstruction techniques, to measure the intensity distributions of X-ray sources of radiotherapy devices. Using this method, we have measured the intensity distributions of X-ray sources from 60Co,6-, 18-, and 25-MV radiotherapy devices. The X-ray sources of the accelerators were all elliptical in shape, but varied in eccentricity, and the sizes of the accelerator sources varied from 0.7 to 3.3 mm full width at half-maximum. The 60Co source was circular in shape and 20 mm in diameter, however, the output from this source was not uniform across its face. The modulation transfer functions (MTF''s) (at the image plane) calculated for the accelerator sources, assuming an image magnification of 1.2, had similar magnitudes at low spatial frequencies as the MTF''s of the metal plate/film detectors commonly used for therapy imaging. However, the source MTF''s declined much more rapidly at high spatial frequencies. Therefore, for geometries commonly found in radiotherapy, the loss in spatial resolution due to the X-ray source was at least equal to that caused by electron and photon scatter within the metal plate/film detectors.Keywords
This publication has 8 references indexed in Scilit:
- Therapy imaging: A signal-to-noise analysis of metal plate/film detectorsMedical Physics, 1987
- A quantitative assessment of portal film contrast as a function of beam energyMedical Physics, 1986
- Digital Imaging For Radiation Therapy VerificationOptical Engineering, 1982
- The significance of screen resolution in treatment verificationInternational Journal of Radiation Oncology*Biology*Physics, 1982
- Fluoroscopic visualization of megavoltage therapeutic x ray beamsInternational Journal of Radiation Oncology*Biology*Physics, 1980
- Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imagingPhysics in Medicine & Biology, 1976
- The value of frequent treatment verification films in reducing localization error in the irradiation of complex fieldsCancer, 1976
- Strip Integration in Radio AstronomyAustralian Journal of Physics, 1956