Comparison of Straight and Curled Tenckhoff Peritoneal Dialysis Catheters Implanted by Percutaneous Technique: A Prospective Randomized Study
Open Access
- 1 January 1995
- journal article
- research article
- Published by SAGE Publications in Peritoneal Dialysis International
- Vol. 15 (1) , 18-21
- https://doi.org/10.1177/089686089501500104
Abstract
Objective: To examine the impact of peritoneal dialysis catheter configuration, curled or straight catheter, on catheter survival and mechanical and infectious complications. Design: Prospective randomized trial. Setting: Department of Nephrology of a single university hospital. Patients: Seventy-two consecutive patients initiating peritoneal dialysis were randomized to receive either a single cuff straight catheter or a single cuff curled catheter, implanted by percutaneous technique. Results: Significantly higher (p < 0.01) survival rate of the curled as compared to the straight catheter. The difference in catheter survival was due to a significantly higher (p < 0.01) incidence of drainage failure associated with catheter tip migration of the straight catheter than of the curled catheter. No difference in infectious complication between the two types of catheters was seen. Catheter survival at 12 months was 77% for the curled catheter and 36% for the straight catheter. Conclusion: This study demonstrates superiority of the curled Tenckhoff peritoneal dialysis catheter survival as compared to the straight catheter. This difference in catheter survival is due to the higher displacement rate of the straight catheter.Keywords
This publication has 9 references indexed in Scilit:
- A Consecutive Study of 646 Peritoneal Dialysis CathetersPeritoneal Dialysis International, 1994
- Catheter Configuration and Outcome in Patients on Continuous Ambulatory Peritoneal Dialysis: A Prospective Comparison of Two CathetersPeritoneal Dialysis International, 1994
- Peritoneal Catheters and Exit -Site Practices: Toward Optimum Peritoneal AccessPeritoneal Dialysis International, 1993
- Six -Year Experience with Swan Neck CathetersPeritoneal Dialysis International, 1992
- The Curled Catheter: Dependable Device for Percutaneous Peritoneal AccessPeritoneal Dialysis International, 1990
- Computerized Tomography With and Without Intraperitoneal Contrast for Determination of Intraabdominal Fluid Distribution and Diagnosis of Complications in Peritoneal Dialysis PatientsAnnual Northeast Bioengineering Conference, 1990
- Chronic Peritoneal Dialysis Catheters: Effects of Catheter Design, Materials, and LocationSeminars in Dialysis, 1990
- The Need for a “Swan Neck” Permanentl Y Bent, Arcuate Peritoneal Dialysis CatheterPeritoneal Dialysis International, 1985