Influence of Flow Rate on Aerosol Particle Size Distributions from Pressurized and Breath-Actuated Inhalers
- 1 January 1998
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine
- Vol. 11 (4) , 231-245
- https://doi.org/10.1089/jam.1998.11.231
Abstract
Particle size distribution of delivered aerosols and the total mass of drug delivered from the inhaler are important determinants of pulmonary deposition and response to inhalation therapy. Inhalation flow rate may vary between patients and from dose to dose. The Andersen Sampler (AS) cascade impactor operated at flow rates of 30 and 55 L/min and the Marple-Miller Impactor (MMI) operated at flow rates of 30, 55, and 80 L/min were used in this study to investigate the influence of airflow rate on the particle size distributions of inhalation products. Total mass of drug delivered from the inhaler, fine particle mass, fine particle fraction, percentage of nonrespirable particles, and amount of formulation retained within the inhaler were determined by ultraviolet spectrophotometry for several commercial bronchodilator products purchased in the marketplace, including a pressurized metered-dose inhaler (pMDI), breath-actuated pressurized inhaler (BAMDI), and three dry powder inhalers (DPIs), two containing salbutamol sulphate and the other containing terbutaline sulphate. Varying the flow rate through the cascade impactor produced no significant change in performance of the pressurized inhalers. Increasing the flow rate produced a greater mass of drug delivered and an increase in respirable particle mass and fraction from all DPIs tested.Keywords
This publication has 33 references indexed in Scilit:
- Comparative Evaluation of Therapeutic Response to Antiasthma Drugs: Dry Powder vs. Conventional Aerosol Delivery SystemsJournal of Aerosol Medicine, 1996
- Asthma Therapy with Aerosols: Clinical Relevance for the Next DecadeJournal of Aerosol Medicine, 1996
- The Montreal Protocol and Essential Use ExemptionsJournal of Aerosol Medicine, 1995
- Comparison of terbutaline and placebo from a pressurised metered dose inhaler and a dry powder inhaler in a subgroup of patients with asthma.Thorax, 1994
- The Evolution of Inhaler TechnologyJournal of Asthma, 1994
- Drug delivery to the respiratory tract using dry powder inhalersInternational Journal of Pharmaceutics, 1994
- A device for overcoming discoordination with metered-dose inhalersJournal of Allergy and Clinical Immunology, 1993
- Lung deposition patterns of directly labelled salbutamol in normal subjects and in patients with reversible airflow obstruction.Thorax, 1993
- Delivery Characteristics of Albuterol Powder Aerosol by RotahalerJournal of Aerosol Medicine, 1993
- Improvement of drug delivery with a breath actuated pressurised aerosol for patients with poor inhaler technique.Thorax, 1991