Preliminary Investigation of Electrothermal Vaporization Sample Introduction for Inductively Coupled Plasma Time-of-Flight Mass Spectrometry
- 20 February 1999
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 71 (7) , 1378-1383
- https://doi.org/10.1021/ac9811625
Abstract
The coupling of an electrothermal vaporization (ETV) apparatus to an inductively coupled plasma time-of-flight mass spectrometer (ICP-TOFMS) is described. The ability of the ICP-TOFMS to produce complete elemental mass spectra at high repetition rates is experimentally demonstrated. A signal-averaging data acquisition board is employed to rapidly record complete elemental spectra throughout the vaporization stage of the ETV temperature cycle; a solution containing 34 elements is analyzed. The reduction of both molecular and atomic isobaric interferences through the temperature program of the furnace is demonstrated. Isobaric overlaps among the isotopes of cadmium, tin, and indium are resolved by exploiting differences in the vaporization characteristics of the elements. Figures of merit for the system are defined with several different data acquisition schemes capable of operating at the high repetition rate of the TOF instrument. With the use of both ion counting and a boxcar averager, the dynamic range is shown to be linear over a range of at least 6 orders of magnitude. A pair of boxcar averagers are used to measure the isotope ratio for silver with a precision of 1.9% RSD, despite a cycle-to-cycle precision of 19% RSD. Detection limits of 10−80 fg are calculated for seven elements, based upon a 10-μL injection.Keywords
This publication has 37 references indexed in Scilit:
- Vaporization and atomization of the platinum group elements in the graphite furnace investigated by electrothermal vaporization-inductively coupled plasma-mass spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1997
- High-Resolution Inductively Coupled Plasma Fourier Transform Ion Cyclotron Resonance Mass SpectrometryAnalytical Chemistry, 1997
- Determination of Inorganic and Total Mercury in Biological Tissues by Electrothermal Vaporization Inductively Coupled Plasma Mass SpectrometryThe Analyst, 1997
- Simultaneous measurement of isotope ratios in solids by laser ablation with a twin quadrupole inductively coupled plasma mass spectrometerJournal of Analytical Atomic Spectrometry, 1995
- Simultaneous Measurement of Ion Ratios by Inductively Coupled Plasma-Mass Spectrometry with a Twin-Quadrupole InstrumentApplied Spectroscopy, 1994
- Performance of an inductively coupled plasma source ion trap mass spectrometerJournal of Analytical Atomic Spectrometry, 1994
- Determination of cadmium and zinc isotope ratios in sheep's blood and organ tissue by electrothermal vaporization inductively coupled plasma mass spectrometry. Invited lectureJournal of Analytical Atomic Spectrometry, 1994
- Electrothermal Vaporization for Sample Introduction in Plasma Source SpectrometryCritical Reviews in Analytical Chemistry, 1992
- Determination of platinum, palladium, ruthenium and iridium geological materials by inductively coupled plasma mass spectrometry with sample introduction by electrothermal vaporisationJournal of Analytical Atomic Spectrometry, 1988
- Inductively coupled plasma source mass spectrometry using continuum flow ion extractionThe Analyst, 1983