Commercial latex agglutination tests for detection of Haemophilus influenzae type b and Streptococcus pneumoniae antigens in patients with bacteremic pneumonia

Abstract
The validity of commercial latex agglutination kits for detection of Haemophilus influenza type b and Streptococcus pneumoniae antigens in serum and urine specimens was studied. We tested serum and urine specimens from 44 patients with bacteremic pneumonia (23 S. pneumoniae, 13 H. influenzae type b, 11 other) with commercial latex agglutination kits (Directigen, Bactigen) for S. pneumoniae and H. influenzae type b antigens. All specimen samples were randomized and read blindly by two readers. Interreader reproducibility was 100%. The sensitivity and specificity of both kits for H. influenzae type b antigens in serum and urine were >90%. None of the 24 urine samples from S. pneumoniae bacteremic patients were positive by either kit, although 76 ng of type 3 polysaccharide could be detected in spiked urine. Sensitivity for S. pneumoniae antigens in serum was 27% for Directigen and 38% for Bactigen. Specificity for S. pneumoniae antigens in serum was 95% for Directigen and 74% for Bactigen. The results suggest that the kits are useful in diagnosing H. influenza type b pneumonia. However, the commercially available S. pneumoniae reagents tested appear to have limited utility for diagnosing S. pneumoniae pneumonia because both kits lack sensitivity and Bactigen lacks specificity, as well.

This publication has 20 references indexed in Scilit: