Abstract
The aim of this study was to investigate cuff compliance and cuff pressure during nitrous oxide exposure in the recently introduced Microcuff tracheal tube with a polyurethane cuff (Microcuff GmbH, Weinheim, Germany), and to compare it to conventional tracheal tubes with PVC cuffs. In an in vitro set up, five cuffed tracheal tubes (ID 7.0 mm) from different manufacturers (Microcuff HVLP, Portex Profile Soft Seal, Mallinckrodt HiLo, Rüsch Super Safety Clear and Sheridan CF) were studied. Pressure-volume curves were assessed and changes of cuff pressure during exposure to nitrous oxide (for 60 min; 66% N(2)O in oxygen) were recorded without and with restriction of the cuff in a trachea model. Each experiment was performed four times using two exemplars of each tube twice. Sixty-minute values of the Microcuff group were compared with the other groups using the Mann- Whitney test (P < 0.05). The Microcuff polyurethane cuff demonstrated an intermediate cuff compliance but the highest cuff pressure increase during nitrous oxide exposure under unrestricted conditions. When inflated within the artificial trachea its cuff compliance became the highest of all tested tracheal tubes. However, exposure to N(2)O, again led to a rapid increase in cuff pressure. The ultra-thin polyurethane tube cuff demonstrated higher permeability for nitrous oxide than conventional PVC cuffs and led to a rapid cuff pressure increase when exposed to N(2)O. Routinely checking of cuff-pressure or filling the cuff with nitrous oxide are more important than the brand of tube used.Keywords
This publication has 16 references indexed in Scilit:
- Air leakage around endotracheal tube cuffsEuropean Journal of Anaesthesiology, 2004
- Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tubeIntensive Care Medicine, 2003
- Time required to achieve a stable cuff pressure by repeated aspiration of the cuff during anaesthesia with nitrous oxideEuropean Journal of Anaesthesiology, 2003
- Changes in gas concentrations in the Brandt endotracheal tube cuff during and after anaesthesia with nitrous oxideEuropean Journal of Anaesthesiology, 2003
- A New Endotracheal Tube with a Cuff Impervious to Nitrous OxideAnesthesia & Analgesia, 1995
- Pressure changes in tracheal tube cuffsAnaesthesia, 1986
- Tracheal tube cuff pressuresAnaesthesia, 1985
- NITROUS OXIDE IN OXYGEN AND TRACHEAL TUBE CUFF VOLUMESBritish Journal of Anaesthesia, 1982
- Physical Characteristics of and Rates of Nitrous Oxide Diffusion into Tracheal Tube CuffsAnesthesiology, 1978