Radiation Exposure to Children in Diagnosing and at Hydrostatic Reduction of Intussusception
- 1 May 1994
- journal article
- research article
- Published by SAGE Publications in Acta Radiologica
- Vol. 35 (3) , 296-299
- https://doi.org/10.1177/028418519403500320
Abstract
The energy imparted to the children in diagnosing and hydrostatic reduction of intussusception was measured in 45 children by means of an area-dose measurement device and the mean absorbed dose was estimated. The device was provided with data on tube kVp, mAs and shutter positions and the results were presented as dGy × cm2. The device had been calibrated against a 30 cm3 ionisation chamber at the relevant kVp range. The median energy imparted and mean absorbed dose were 10.8 mJ and 0.94 mGy, respectively. 70% of the total dose was delivered during fluoroscopy. The complex irradiation situation with varying field collimation, tube voltage and amount of photon absorbing barium sulphate in the intestines renders organ dose and hazards estimations less reliable. However, even leaving the radiation shielding effect of the barium sulphate out, the radiation load is justifiable for a combined diagnostic and interventional procedure.Keywords
This publication has 11 references indexed in Scilit:
- Conversion Factors between Energy Imparted to the Patient and Air Collision Kerma Integrated over Beam Area in Pediatric RadiologyActa Radiologica, 1993
- Relations between effective dose equivalent and mean absorbed dose (energy imparted) to patients in diagnostic radiologyPhysics in Medicine & Biology, 1986
- Doses to patients from routine diagnostic X-ray examinations in EnglandThe British Journal of Radiology, 1986
- Energy imparted to the patient in diagnostic radiology: calculation of conversion factors for determining the energy imparted from measurements of the air collision kerma integrated over beam areaPhysics in Medicine & Biology, 1984
- The measurement of energy imparted to patients during diagnostic x-ray examinations using the Diamentor exposure-area product meterPhysics in Medicine & Biology, 1984
- An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinationsPhysics in Medicine & Biology, 1982
- Radiation Doses to Children at Urologic RadiographyActa Radiologica. Diagnosis, 1981
- Energy Imparted in Roentgen Diagnostic ProceduresActa Radiologica. Diagnosis, 1979
- Determination of Integral Absorbed Dose from Exposure MeasurementsActa Radiologica: Therapy, Physics, Biology, 1963