Novel Immunofluorescence Assay Using Recombinant Nucleocapsid-Spike Fusion Protein as Antigen To Detect Antibodies against Severe Acute Respiratory Syndrome Coronavirus
Open Access
- 1 February 2005
- journal article
- Published by American Society for Microbiology in Clinical and Vaccine Immunology
- Vol. 12 (2) , 321-328
- https://doi.org/10.1128/cdli.12.2.321-328.2005
Abstract
Severe acute respiratory syndrome (SARS) is caused by a novel and highly infectious virus named SARS coronavirus (SARS-CoV). Among the serological tests currently available for the detection of SARS-CoV, a whole-virus-based immunofluorescence assay (IFA) was considered one of the most sensitive assays and served as a “gold standard” during the SARS epidemic in Singapore in 2003. However, the need to manipulate live SARS-CoV in the traditional IFA limits its wide application due to the requirement for a biosafety level 3 laboratory and the risk of laboratory infection. Previously, we have identified two immunodominant epitopes, named N195 and Sc, in the two major structural proteins, the N and S proteins, of SARS-CoV (Q. He, K. H. Chong, H. H. Chng, B. Leung, A. E. Ling, T. Wei, S. W. Chan, E. E. Ooi, and J. Kwang, Clin. Diagn. Lab. Immunol., 11:417-422, 2004; L. Lu, I. Manopo, B. P. Leung, H. H. Chng, A. E. Ling, L. L. Chee, E. E. Ooi, S. W. Chan, and J. Kwang, J. Clin. Microbiol. 42:1570-1576, 2004). In the present study, the N195-Sc fusion protein was highly expressed in insect (Sf9) cells infected with a recombinant baculovirus bearing the hybrid gene under the control of a polyhedrin promoter. An IFA based on Sf9 cells producing the fusion protein was standardized with 23 serum samples from patients with SARS, 20 serum samples from patients with autoimmune diseases, and 43 serum samples from healthy blood donors. The detection rates were comparable to those obtained with a commercial SARS-CoV IFA kit (EUROIMMUN, Gross Groenau, Germany) and a conventional IFA performed at the Singapore General Hospital. Our data showed that the newly developed IFA could detect SARS-CoV in 22 of the 23 SARS-CoV-positive serum samples and gave no false-positive results when the sera from patients with autoimmune diseases and healthy individuals were tested. The detection rate was identical to those of the two whole-virus-based IFAs. Thus, the novel N-S fusion antigen-based IFA could be an attractive alternative to present whole-virus-based IFAs for the diagnosis of SARS-CoV infection.Keywords
This publication has 9 references indexed in Scilit:
- Immunological Characterization of the Spike Protein of the Severe Acute Respiratory Syndrome CoronavirusJournal of Clinical Microbiology, 2004
- Development of a Western Blot Assay for Detection of Antibodies against Coronavirus Causing Severe Acute Respiratory SyndromeClinical and Vaccine Immunology, 2004
- Maintenance of Insect Cell Cultures and Generation of Recombinant BaculovirusesCurrent Protocols in Molecular Biology, 2004
- Isolation and Characterization of Viruses Related to the SARS Coronavirus from Animals in Southern ChinaScience, 2003
- Evaluation of Reverse Transcription-PCR Assays for Rapid Diagnosis of Severe Acute Respiratory Syndrome Associated with a Novel CoronavirusJournal of Clinical Microbiology, 2003
- A Novel Coronavirus Associated with Severe Acute Respiratory SyndromeNew England Journal of Medicine, 2003
- Identification of a Novel Coronavirus in Patients with Severe Acute Respiratory SyndromeNew England Journal of Medicine, 2003
- Immunofluorescence Technique Using HeLa Cells Expressing Recombinant Nucleoprotein for Detection of Immunoglobulin G Antibodies to Crimean-Congo Hemorrhagic Fever VirusJournal of Clinical Microbiology, 2002
- Immunofluorescence Method for Detection of Ebola Virus Immunoglobulin G, Using HeLa Cells Which Express Recombinant NucleoproteinJournal of Clinical Microbiology, 2001