Cardiac magnetic resonance parallel imaging at 3.0 Tesla: Technical feasibility and advantages
- 24 February 2004
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 19 (3) , 291-297
- https://doi.org/10.1002/jmri.20015
Abstract
Purpose: To quantify changes in signal‐to‐noise ratio (SNR), contrast‐to‐noise ratio (CNR), specific absorption rate (SAR), RF power deposition, and imaging time in cardiac magnetic resonance imaging with and without the application of parallel imaging at 1.5 T and 3.0 T.Materials and Methods: Phantom and volunteer data were acquired at 1.5 T and 3.0 T with and without parallel imaging.Results: Doubling field strength increased phantom SNR by a factor of 1.83. In volunteer data, SNR and CNR values increased by factors of 1.86 and 1.35, respectively. Parallel imaging (reduction factor = 2) decreased phantom SNR by a factor of 1.84 and 2.07 when compared to the full acquisition at 1.5 T and 3.0 T, respectively. In volunteers, SNR and CNR decreased by factors of 2.65 and 2.05 at 1.5 T and 1.99 and 1.75 at 3.0 T, respectively. Doubling the field strength produces a nine‐fold increase in SAR (0.0751 to 0.674 W/kg). Parallel imaging reduced the total RF power deposition by a factor of two at both field strengths.Conclusions: Parallel imaging decreases total scan time at the expense of SNR and CNR. These losses are compensated at higher field strengths. Parallel imaging is effective at reducing total power deposition by reducing total scan time. J. Magn. Reson. Imaging 2004;19:291–297.Keywords
This publication has 12 references indexed in Scilit:
- Cardiac Function: MR Evaluation in One Breath Hold with Real-time True Fast Imaging with Steady-State PrecessionRadiology, 2002
- Specific coil design for SENSE: A six-element cardiac arrayMagnetic Resonance in Medicine, 2001
- Sensitivity and Power Deposition in a High-Field Imaging ExperimentJournal of Magnetic Resonance Imaging, 2000
- Cardiac real-time imaging using SENSEMagnetic Resonance in Medicine, 2000
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- The Intrinsic Signal-to-Noise Ratio in Human Cardiac Imaging at 1.5, 3, and 4 TJournal of Magnetic Resonance, 1997
- MR desktop dataJournal of Magnetic Resonance Imaging, 1992
- Volume imaging with MR phased arraysMagnetic Resonance in Medicine, 1991
- The NMR phased arrayMagnetic Resonance in Medicine, 1990
- The field dependence of NMR imaging. II. Arguments concerning an optimal field strengthMagnetic Resonance in Medicine, 1986