Magnetic Resonance Monitoring of Stent Deployment
- 1 June 2000
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 35 (6) , 343-351
- https://doi.org/10.1097/00004424-200006000-00002
Abstract
To evaluate MR imaging features of commercially available stents before, during, and after in vitro deployment as a step toward MR-guided stent deployment. Fourteen stents were deployed in a phantom under MR monitoring at 1.5 T by using a gradient-echo sequence. Device visibility was rated on a four-point scale (excellent, fair, poor, not visible). The Memotherm stent and the rolling membrane (RM) Wallstent showed excellent stent visibility and at least fair scores for artifact-induced narrowing of the stent lumen. Three stents (Palmaz, AVE, Easy Wallstent) showed excellent visibility of the stent but no visible lumen. Five stents (Strecker, Accuflex, Hemobahn, Passager, Sinus) displayed fair visibility. The delivery catheters of four stent systems (Smart, Vascucoil, Symphony, ZA) displayed severe black hole artifacts. The imaging features of several stent systems might be suitable for MR-guided intervention. The Memotherm and the Wallstent RM combine good visibility of the stent and the lumen.Keywords
This publication has 26 references indexed in Scilit:
- MR-guided Transjugular Portosystemic Shunt Placement in a Swine ModelJournal of Vascular and Interventional Radiology, 1999
- MR‐guided placement of a temporary vena cava filter: Technique and feasibilityJournal of Magnetic Resonance Imaging, 1998
- MR‐guided balloon angioplasty: In vitro demonstration of the potential of MRI for guiding, monitoring, and evaluating endovascular interventionsJournal of Magnetic Resonance Imaging, 1998
- MR imaging-guided intravascular procedures: initial demonstration in a pig model.Radiology, 1997
- Will It Be Feasible to Insert Endoprostheses Under Interventional MRI?Journal of Endovascular Surgery, 1996
- Measurement of the arterial concentration of Gd‐DTPA using MRI: A step toward quantitative perfusion imagingMagnetic Resonance in Medicine, 1996
- Vascular interventions guided by ultrafast MR imaging: Evaluation of different materialsMagnetic Resonance in Medicine, 1994
- Real‐time position monitoring of invasive devices using magnetic resonanceMagnetic Resonance in Medicine, 1993
- Shunts for portal hypertension: MR and angiography for determination of patency.Radiology, 1986
- Susceptibility artefacts in NMR imagingMagnetic Resonance Imaging, 1985