Variations in Sample pH and pO2 Affect ExacTech Meter Glucose Measurements
- 1 June 1994
- journal article
- Published by Wiley in Diabetic Medicine
- Vol. 11 (5) , 506-509
- https://doi.org/10.1111/j.1464-5491.1994.tb00315.x
Abstract
A study was conducted to assess the effect of changes in sample pH and pO2 on the Accutrend, ExacTech Companion, Glucometer II, One Touch II, and Refloflux II blood glucose meters. Venous blood was tonometered after the addition of strong acid or alkali to produce 10 samples with a pH range of 6.54 to 7.73 but with a similar mean pO2 (9.09 ± 0.096 (SEM) kPa) and pCO2 (5.52 ± 0.024 kPa). The ExacTech showed a > 15% deviation from the value obtained at pH 7.40 below pH 6.95 and above pH 7.85 (r = 0.69, p = 0.026). Ten further samples were made with a pO2 range of 2.0 to 33.6 kPa but mean pH (7.28 ± 0.012) and pCO2 (7.98 ± 0.25 kPa) kept alike. The ExacTech was accurate at pO2=11.5 kPa but deviated by > 15% below pO2 = 5.8 kPa and above 22.9 kPa (r = −0.95, p = 0.00002). The Accutrend and One Touch II required a PO2 change of 42 and 45 kPa to achieve the same error (r = (r 0.82 and r = −0.67 respectively). Thus caution is required in the interpretation of ExacTech glucose measurements in severely acidotic or hypoxic patients.Keywords
This publication has 12 references indexed in Scilit:
- Intra‐operative blood glucose measurementsAnaesthesia, 1994
- The Effect of Variations in Haematocrit, Mean Cell Volume and Red Blood Cell Count on Reagent Strip Tests for GlucoseAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1993
- Capillary and Venous Blood Glucose Measurements Using a Direct Glucose‐sensing MeterDiabetic Medicine, 1991
- The Effect of Haematocrit on Reagent Strip Tests for GlucoseDiabetic Medicine, 1991
- Effect of temperature on glucose results obtained with the Reflolux IIClinical Chemistry, 1990
- Influence of blood oxygen tension on dipstick glucose determinations.Clinical Chemistry, 1989
- Colour vision and blood glucose self-monitoring in diabeticsPractical Diabetes International, 1988
- Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer IIDiabetes Care, 1988
- PEN-SIZED DIGITAL 30-SECOND BLOOD GLUCOSE METERThe Lancet, 1987
- Ferrocene-mediated enzyme electrode for amperometric determination of glucoseAnalytical Chemistry, 1984