Quantitative estimation of quaternary ammonium neuromuscular blocking agents in serum by direct insertion probe chemical ionization mass spectrometry
- 1 July 1986
- journal article
- research article
- Published by Wiley in Journal of Mass Spectrometry
- Vol. 13 (7) , 327-332
- https://doi.org/10.1002/bms.1200130703
Abstract
A new method based on selected ion monitoring chemical ionization mass spectrometry was developed to measure the quaternary ammonium neuromuscular blocking agents, pancuronium and vecuronium, in serum. An ion‐pair extraction procedure is utilized to separate the compounds of interest from biological fluids. The intensities of the ion currents produced by the bisamines formed from the drugs and the corresponding deuterated internal standards through thermolytic dequaternization are monitored. The assay shows good linearity over the range of 1–500 ng/ml. This assay has been utilized in a variety of clinical pharmacokinetic studies involving surgical pediatric, geriatric and obstetric patients requiring anesthesia.Keywords
This publication has 15 references indexed in Scilit:
- Clinical Pharmacology of ORG NC45 (NorcuronTM)Anesthesiology, 1981
- A Comparison of the Neuromuscular Blocking and Vagolytic Effects of ORG NC45 and PancuroniumAnesthesiology, 1981
- Determination of Org NC 45 (a myoneural blocking agent) in human plasma using high-performance normal-phase liquid chromatographyJournal of Chromatography B: Biomedical Sciences and Applications, 1980
- Pharmacokinetics and Pharmacodynamics of ORG NC 45 in ManBritish Journal of Anaesthesia, 1980
- The neuromuscular and autonomic blocking activities of pancuronium, Org NC 45, and other pancuronium analogues, in the catJournal of Pharmacy and Pharmacology, 1979
- Modified Fluorometric Quantitation of Pancuronium Bromide and Metabolites in Human Maternal and Umbilical SerumsJournal of Pharmaceutical Sciences, 1979
- A modified method for the fluorimetric determination of pancuronium bromide in plasmaClinica Chimica Acta; International Journal of Clinical Chemistry, 1977
- Determination of pancuronium bromide and its metabolites in human urine by dye-extraction method - relation between the extractability and structure of quaternary ammonium ions.CHEMICAL & PHARMACEUTICAL BULLETIN, 1974
- Fluorimetric and chromatographic determination of pancuronium bromide and its metabolites in biological materialsClinica Chimica Acta; International Journal of Clinical Chemistry, 1973
- Pancuronium BromideDrugs, 1972