Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer
Top Cited Papers
- 18 February 2006
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 78 (7) , 2113-2120
- https://doi.org/10.1021/ac0518811
Abstract
Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60 000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100 000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.Keywords
This publication has 9 references indexed in Scilit:
- Interfacing the Orbitrap Mass Analyzer to an Electrospray Ion SourceAnalytical Chemistry, 2003
- A two-dimensional quadrupole ion trap mass spectrometerPublished by American Chemical Society (ACS) ,2002
- An introduction to quadrupole–time‐of‐flight mass spectrometryJournal of Mass Spectrometry, 2001
- Electrostatic Axially Harmonic Orbital Trapping: A High-Performance Technique of Mass AnalysisAnalytical Chemistry, 2000
- Quantitation of ion abundances in fourier transform ion cyclotron resonance mass spectrometryJournal of the American Society for Mass Spectrometry, 1998
- External accumulation of ions for enhanced electrospray ionization fourier transform ion cyclotron resonance mass spectrometryJournal of the American Society for Mass Spectrometry, 1997
- High Sensitivity Collisionally-activated Decomposition Tandem Mass Spectrometry on a Novel Quadrupole/Orthogonal-acceleration Time-of-flight Mass SpectrometerRapid Communications in Mass Spectrometry, 1996
- Experimental determination of the number of trapped ions, detection limit, and dynamic range in Fourier transform ion cyclotron resonance mass spectrometryAnalytical Chemistry, 1993
- A Method for the Neutralization of Electron Space Charge by Positive Ionization at Very Low Gas PressuresPhysical Review B, 1923