Abstract
Esophageal Temperature Monitoring During PVAC Ablation. Introduction: A novel ablation system has been introduced for rapid treatment of atrial fibrillation (AF). This system delivers duty‐cycled phased radiofrequency (RF) energy via an over‐the‐wire catheter (PVAC®, Medtronic) to achieve pulmonary vein (PV) isolation. Lower power and depth control suggests that collateral damage might be minimized. However, no studies have investigated the potential for thermal effect and damage to the esophagus. Methods and Results: Ninety consecutive patients undergoing PV‐isolation were evaluated. Group A (48 patients) had continuous luminal esophageal temperature (LET) monitoring using a temperature probe with 3 metal electrodes located in the vicinity of the targeted PV ostia. Ablation ceased when LET exceeded 40 °C. Only patients with LET ≥ 39 °C underwent endoscopic evaluation to assess esophageal damage. Group B (42 patients) excluded LET monitoring but all patients underwent endoscopy. In Group A, 27 (56%) patients showed LET ≥ 39 °C (mean LET 40.5 °C). Endoscopy revealed esophageal alterations in 5 (8%) (3 erythema and 2 intramural bleeding). One hundred eighty‐nine out of 190 (99.5%) targeted PVs were successfully isolated, with 1 PV unsuccessful due to high LET. In Group B all 165 targeted PVs (100%) were successfully isolated. Endoscopy in Group 2 revealed no esophageal alterations. Conclusions: Using a duty‐cycled, phased RF ablation system is safe and effective to isolate PVs. No Eso alteration was documented after ablation when LET was not monitored. This suggests that the LET probe may contribute to the thermal effect. Whether the documented increments in LET are due to direct tissue heating or possible interaction between the LET probe requires further investigation. (J Cardiovasc Electrophysiol, Vol. 22, pp. 255‐261, March 2011)Keywords
This publication has 28 references indexed in Scilit:
- Catheter Ablation for Paroxysmal Atrial Fibrillation: A Randomized Comparison between Multielectrode Catheter and Point-by-Point AblationPacing and Clinical Electrophysiology, 2010
- Pulmonary vein isolation using a novel decapolar over-the-wire mapping and ablation catheterExpert Review of Cardiovascular Therapy, 2009
- Characterization of In Vitro and In Vivo Lesions Made by a Novel Multichannel Ablation Generator and a Circumlinear Decapolar Ablation CatheterJournal of Cardiovascular Electrophysiology, 2009
- Treatment of Atrial Fibrillation With Antiarrhythmic Drugs or Radiofrequency AblationCirculation: Arrhythmia and Electrophysiology, 2009
- Treatment of Cardiac Tamponade During Electrophysiology Procedures: Do Electrophysiologists Need More Than Drainage?Journal of Cardiovascular Electrophysiology, 2009
- Pulmonary vein isolation by duty-cycled bipolar and unipolar radiofrequency energy with a multielectrode ablation catheterHeart Rhythm, 2008
- A nationwide survey on the prevalence of atrioesophageal fistula after left atrial radiofrequency catheter ablationJournal of Interventional Cardiac Electrophysiology, 2008
- Topographic Variability of the Esophageal Left Atrial Relation Influencing Ablation Lines in Patients with Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2005
- Left Atrial–Esophageal Fistula Following Radiofrequency Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2004
- Atrio-Esophageal Fistula as a Complication of Percutaneous Transcatheter Ablation of Atrial FibrillationCirculation, 2004