Performance of QPET, a high-resolution 3D PET imaging system for small volumes
- 1 March 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 13 (1) , 176-185
- https://doi.org/10.1109/42.276156
Abstract
QPET is a positron imaging system developed at Queen's University for high-resolution, 3D imaging of small volumes; it includes a pair of planar gamma-ray detectors 25.4 cm square, which rotate about a central axis with a quasi-cylindrical geometry. The authors describe the performance of this system. Basic characteristics of the detectors are evaluated: a spatial sampling of 1 mm, a quantum efficiency of 9.3% (for 511 keV gamma rays with normal incidence), and a time resolution of 88 ns. Models are developed to characterize the system deadtime and the sensitivity in terms of the noise-equivalent counting rate. With an 8 cm diameter spherical source, the noise-equivalent counting rate reaches a maximum at just over 3 kcps for an activity concentration of 2 muCi/cc; the random coincidence events and the deadtime losses both contribute significantly and the scatter contribution is small. Spatial resolution and uniformity over the field of view are evaluated by imaging short and long line sources; a spatial resolution of 2.7 mm in the transverse directions and 2.0 mm in the axial direction is achieved, with excellent uniformity throughout the field of view. The detector response is amplitude invariant across a 20 cm transverse diameter and a 9 cm axial length with the acceptance angle limited to +/-25 degrees in the axial direction. As an example of the imaging capabilities of QPET, the authors show 3D images of (18)F uptake in the bones of a rat, showing the excellent spatial resolution. This system is best suited to limited-volume applications where high counting rates are not necessary, but where high spatial resolution and uniform detector response are priorities.Keywords
This publication has 30 references indexed in Scilit:
- 2D and 3D imaging of small animals and the human radial artery with a high resolution detector for PETIEEE Transactions on Medical Imaging, 1992
- A deconvolution scatter correction for a 3-D PET systemIEEE Transactions on Medical Imaging, 1992
- Object shape dependent scatter simulations for PETIEEE Transactions on Nuclear Science, 1991
- Dynamic high resolution imaging of rats: design considerationsIEEE Transactions on Nuclear Science, 1991
- Image reconstruction for a 3D PET system using a minimum norm constraintPhysics in Medicine & Biology, 1991
- Detector, shielding and geometric design factors for a high-resolution PET systemIEEE Transactions on Nuclear Science, 1990
- PET system calibrations and corrections for quantitative and spatially accurate imagesIEEE Transactions on Nuclear Science, 1989
- Compensation for crystal penetration in high resolution positron tomographyIEEE Transactions on Nuclear Science, 1989
- Compton scattering in a large-aperture positron imaging systemIEEE Transactions on Medical Imaging, 1988