Use of conventional ECG electrodes for depth of anaesthesia monitoring using the cerebral state index: a clinical study in day surgery
Open Access
- 19 March 2007
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 98 (5) , 645-648
- https://doi.org/10.1093/bja/aem061
Abstract
The cost–benefit relationship for depth of anaesthesia monitors is complicated by the high cost of specially designed EEG electrodes. The cerebral state index (CSI) monitor will accept regular ECG electrodes with snap connectors. The purpose of this study was to determine if generic ECG electrodes could replace the more expensive proprietary EEG electrodes for the CSI monitor. Two identical cerebral state monitors were used simultaneously during sevoflurane anaesthesia for knee arthroscopy in 14 ASA I–II patients. One monitor used proprietary (Danmeter) EEG electrodes and the other used ECG electrodes (3M™ Red Dot™ Diagnostic ECG Electrodes). Paired CSI values were recorded every other minute. Anaesthetic depth was titrated clinically. Sedation depth was scored according to the Observer's Assessment of Alertness/Sedation (OAAS) scale. The agreement between the two measures was found to be high, mean difference − 0.23, and the overall repeatability mean bias was 6.6 and 153/163 pairs (94%) were located within the 95% limits of agreement. No major difference was noted in impedance, noise, or artifacts. A large overlap in CSI was noted for each level of the OAAS scale; patients with CSI values as low as 40–50 responded whereas patients not responding to surgical stimulation had CSI values as high as 75. The direct cost of disposables decreased from 4€ to 0.50€ per patient by using ordinary ECG electrodes. Switching from proprietary EEG electrodes to ordinary generic ECG electrodes maintains the same accuracy at about a 10th of the cost when measuring CSI during day surgery with sevoflurane anaesthesia.Keywords
This publication has 10 references indexed in Scilit:
- Cerebral state monitor, a new small handheld EEG monitor for determining depth of anaesthesiaEuropean Journal of Anaesthesiology, 2006
- Comparative evaluation of the cerebral state index and the bispectral index during target-controlled infusion of propofolBritish Journal of Anaesthesia, 2005
- The Incidence of Awareness During Anesthesia: A Multicenter United States StudyAnesthesia & Analgesia, 2004
- Bispectral index monitoring to prevent awareness during anaesthesia: the B-Aware randomised controlled trialThe Lancet, 2004
- Reduction in the incidence of awareness using BIS monitoringActa Anaesthesiologica Scandinavica, 2003
- Bispectral index monitoring: comparison of two types of electrodeAnaesthesia, 2000
- Comparing Methods of Clinical Measurement: Reporting Standards for Bland and Altman AnalysisAnesthesia & Analgesia, 2000
- EEG bispectral index monitoring in sevoflurane or propofol anaesthesia: analysis of direct costs and immediate recoveryActa Anaesthesiologica Scandinavica, 1999
- Measuring the Performance of Anesthetic Depth IndicatorsAnesthesiology, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986