Effect of a spacer on pulmonary aerosol deposition from a jet nebuliser during mechanical ventilation.
Open Access
- 1 January 1995
- Vol. 50 (1) , 50-53
- https://doi.org/10.1136/thx.50.1.50
Abstract
BACKGROUND--Several factors have been identified which improve nebulised aerosol delivery in vitro. One of these is the addition of a spacer to the ventilator circuit which improves aerosol delivery from a jet nebuliser to a model lung by approximately 30%. The current study was designed to demonstrate whether similar improvements could be demonstrated in vivo. METHODS--Ten patients (seven men) were studied during mechanical ventilation (Siemens Servo 900C) after open heart surgery. Aerosol was delivered using a Siemens Servo 945 nebuliser system (high setting) driving a System 22 Acorn jet nebuliser (Medic-Aid) containing 3 ml technetium-99m labelled human serum albumin (99mTc-HSA (50 micrograms); activity in the first nebulisation, 90 MBq; in the second nebulisation, 185 MBq). Central and peripheral lung aerosol deposition and the time to complete deposition were measured using a gamma camera and compared when the nebuliser was connected to the inspiratory limb using a simple T-piece or a 600 ml spacer. RESULTS--The addition of the spacer increased total lung deposition (mean (SD) percentage initial nebuliser activity) from 2.2 (0.7)% to 3 (0.8)%. There was no difference in the time required to complete nebulisation (18.2 min v 18.3 min respectively for T-piece and spacer) or in the retention of activity in the nebuliser (46.2% v 47.1% respectively). CONCLUSIONS--The combination of a spacer with a jet nebuliser increased lung deposition by 36% in mechanically ventilated patients and is a simple way of increasing drug deposition or reducing the amount of an expensive drug required for nebulisation.Keywords
This publication has 10 references indexed in Scilit:
- Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery.American Journal of Respiratory and Critical Care Medicine, 1994
- Delivery of Ultrasonic Nebulized Aerosols to a Lung Model during Mechanical VentilationAmerican Review of Respiratory Disease, 1993
- Pulmonary deposition of a nebulised aerosol during mechanical ventilation.Thorax, 1993
- Delivery of a Nebulized Aerosol to a Lung Model during Mechanical Ventilation: Effect of Ventilator Settings and Nebulizer Type, Position, and Volume of FillAmerican Review of Respiratory Disease, 1992
- Nebulizer Function during Mechanical VentilationAmerican Review of Respiratory Disease, 1992
- Pressurized Aerosol versus Jet Aerosol Delivery to Mechanically Ventilated Patients Comparison of Dose to the LungsAmerican Review of Respiratory Disease, 1990
- Improving the efficiency of drug administration with jet nebulisers.1988
- Aerosol delivery in intubated, mechanically ventilated patientsCritical Care Medicine, 1985
- Improvement of pressurised aerosol deposition with Nebuhaler spacer device.Thorax, 1984
- DEPOSITION OF PRESSURIZED SUSPENSION AEROSOLS INHALED THROUGH EXTENSION DEVICESPublished by Elsevier ,1981