Attenuation correction of myocardial SPECT images with X-ray CT: Effects of registration errors between X-ray CT and SPECT
- 1 September 2002
- journal article
- research article
- Published by Springer Nature in Annals of Nuclear Medicine
- Vol. 16 (6) , 431-435
- https://doi.org/10.1007/bf02990083
Abstract
Purpose: Attenuation correction with an X-ray CT image is a new method to correct attenuation on SPECT imaging, but the effect of the registration errors between CT and SPECT images is unclear. In this study, we investigated the effects of the registration errors on myocardial SPECT, analyzing data from a phantom and a human volunteer.Methods: Registerion (fusion) of the X-ray CT and SPECT images was done with standard packaged software in three dimensional fashion, by using linked transaxial, coronal and sagittal images. In the phantom study, an X-ray CT image was shifted 1 to 3 pixels on thex, y andz axes, and rotated 6 degrees clockwise. Attenuation correction maps generated from each misaligned X-ray CT image were used to reconstruct misaligned SPECT images of the phantom filled with201Tl. In a human volunteer, X-ray CT was acquired in different conditions (during inspiration vs. expiration). CT values were transferred to an attenuation constant by using straight lines; an attenuation constant of 0/cm in the air (CT value=−1,000 HU) and that of 0.150/cm in water (CT value=0 HU). For comparison, attenuation correction with transmission CT (TCT) data and an external γ-ray source (99mTc) was also applied to reconstruct SPECT images.Results: Simulated breast attenuation with a breast attachment, and inferior wall attenuation were properly corrected by means of the attenuation correction map generated from X-ray CT. As pixel shift increased, deviation of the SPECT images increased in misaligned images in the phantom study. In the human study, SPECT images were affected by the scan conditions of the X-ray CT.Conclusion: Attenuation correction of myocardial SPECT with an X-ray CT image is a simple and potentially beneficial method for clinical use, but accurate registration of the X-ray CT to SPECT image is essential for satisfactory attenuation correction.Keywords
This publication has 10 references indexed in Scilit:
- CT vs 68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube currentEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- Receiver operating characteristic (ROC) analysis of images reconstructed with iterative expectation maximization algorithms.Annals of Nuclear Medicine, 2001
- Phantom study of fusion image of CT and SPECT with body-contour generated from external Compton scatter sources.Annals of Nuclear Medicine, 2000
- Gamma camera-mounted anatomical X-ray tomography: technology, system characteristics and first imagesEuropean Journal of Nuclear Medicine and Molecular Imaging, 2000
- MRI-SPET and SPET-SPET brain co-registrationNuclear Medicine Communications, 1999
- Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimetersEuropean Journal of Nuclear Medicine and Molecular Imaging, 1998
- A Fully Automatic Multimodality Image Registration AlgorithmJournal of Computer Assisted Tomography, 1995
- Improvement of brain single photon emission tomography (SPET) using transmission data acquisition in a four-head SPET scannerEuropean Journal of Nuclear Medicine and Molecular Imaging, 1993
- A technique for using CT images in attenuation correction and quantification in SPECTNuclear Medicine Communications, 1989