Anisotropic noise propagation in diffusion tensor MRI sampling schemes
Open Access
- 16 May 2003
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 49 (6) , 1143-1151
- https://doi.org/10.1002/mrm.10491
Abstract
The subject of this study is the controversial choice of directions in diffusion tensor MRI (DT‐MRI); specifically, the numerical algebra related to this choice. In DT‐MRI, apparent diffusivities are sampled in six or more directions and a least‐squares equation is solved to reconstruct the diffusion tensor. Numerical characteristics of the system are considered, in particular the condition number and normal matrix, and are shown to be dependent on the relative orientation of the tensor with respect to the laboratory frame. As a consequence, noise propagation can be anisotropic. However, the class of icosahedral direction schemes is an exception, and icosahedral directions have the same condition number and normal matrix for direction encoding as the ideal scheme with an infinite number of directions. This normal matrix and its condition number are rotationally invariant. Numerical simulations show that for icosahedral schemes with 30 directions the standard deviation of the fractional anisotropy is both low and nearly independent of fiber orientation. The recommended choice of directions for a DT‐MRI experiment is therefore the icosahedral set of directions with the highest number of directions achievable in the available time. Magn Reson Med 49:1143–1151, 2003.Keywords
This publication has 16 references indexed in Scilit:
- Theoretical analysis of the effects of noise on diffusion tensor imagingMagnetic Resonance in Medicine, 2001
- Comparison of gradient encoding schemes for diffusion‐tensor MRIJournal of Magnetic Resonance Imaging, 2001
- Water diffusion compartmentation and anisotropy at high b values in the human brainMagnetic Resonance in Medicine, 2000
- Condition Number as a Measure of Noise Performance of Diffusion Tensor Data Acquisition Schemes with MRIJournal of Magnetic Resonance, 2000
- Minimal gradient encoding for robust estimation of diffusion anisotropy☆Magnetic Resonance Imaging, 2000
- Optimal strategies for measuring diffusion in anisotropic systems by magnetic resonance imagingMagnetic Resonance in Medicine, 1999
- A Comparative Study of Acquisition Schemes for Diffusion Tensor Imaging Using MRIJournal of Magnetic Resonance, 1999
- Distributing many points on a sphereThe Mathematical Intelligencer, 1997
- Estimation of the Effective Self-Diffusion Tensor from the NMR Spin EchoJournal of Magnetic Resonance, Series B, 1994
- Hecke operators and distributing points on the sphere ICommunications on Pure and Applied Mathematics, 1986