Undesirable Mode Switching with a Dual Chamher Rate Responsive Pacemaker
- 1 April 1993
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (4) , 729-737
- https://doi.org/10.1111/j.1540-8159.1993.tb01652.x
Abstract
The Telectronics 1250 Meta MV DDDR pacemaker is a new device featuring automatic mode switching from DDDR to VVIR pacing in the event of an atrial arrhythmia. Although mode switching is a valuable feature, sinus tachycardia can cause an undesirable mode switch to occur. Of 24 implants at this institution, 11 have been for an AV conduction disorder. Eight of these 11 patients were specifically evaluated for undesirable mode switching. During exercise testing and/or Holter monitoring, mode switching was repeatedly seen in seven of the eight at low levels of exercise. Factors precipitating mode switching were a low rate response factor, low upper rate setting, long base postventricular atrial refractory period (PVARP) and a long AV delay. During Holter monitoring, patients spent up to 50% of the time in VVIR pacing as opposed to DDDR pacing. It is concluded that patients with intact sinus node function are at risk of undesirable mode switching and should probably be programmed to the DDD mode unless there is a specific indication for DDDR pacing. If the DDDR mode is chosen, careful selection of the aforementioned pacing parameters is required.Keywords
This publication has 9 references indexed in Scilit:
- Double blind crossover comparison of the effects of dual chamber pacing (DDD) and ventricular rate adaptive (VVIR) pacing on neuroendocrine variables, exercise performance, and symptoms in complete heart block.Heart, 1991
- A randomized double-blind crossover comparison of four rate-responsive pacing modesJournal of the American College of Cardiology, 1991
- VVIR or DDD(R): Does it matter?Clinical Cardiology, 1991
- Clinical Predictors and Natural History of Atrial Fibrillation in Patients with DDD PacemakersPacing and Clinical Electrophysiology, 1990
- Symptoms, Cardiovascular Risk Profile and Spontaneous ECG in Paced Patients: A Five‐Year Follow‐Up StudyPacing and Clinical Electrophysiology, 1990
- Intrapatient Comparison Between Chronic VVIR and DDD Pacing ‘In Patients Affected by High Degree AV Block Without Heart FailurePacing and Clinical Electrophysiology, 1990
- Pacing for Carotid Sinus Syndrome and Sick Sinus SyndromePacing and Clinical Electrophysiology, 1990
- Rate Responsive Pacing Using a Minute Ventilation SensorPacing and Clinical Electrophysiology, 1988
- Initial Clinical Experience with a Minute Ventilation Sensing Rate Modulated Pacemaker: Improvements in Exercise Capacity and SymptomatologyPacing and Clinical Electrophysiology, 1988