Dosimetric properties of improved GafChromic films for seven different digitizers
Top Cited Papers
- 13 August 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (9) , 2392-2401
- https://doi.org/10.1118/1.1776691
Abstract
Two recently introduced GafChromic film models, HS and XR‐T, have been developed as more sensitive and uniform alternatives to GafChromic MD‐55‐2 film. The HS model has been specifically designed for measurement of absorbed dose in high‐energy photon beams (above 1 MeV), while the XR‐T model has been introduced for dose measurements of low energy (0.1 MeV) photons. The goal of this study is to compare the sensitometric curves and estimated dosimetric uncertainties associated with seven different GafChromic film dosimetry systems for the two new film models. The densitometers tested are: LKB Pharmacia UltroScan XL, Molecular Dynamics Personal Densitometer, Nuclear Associates Radiochromic Densitometer Model 37‐443, Photoelectron Corporation CMR‐604, Laser Pro 16, Vidar VXR‐16, and AGFA Arcus II document scanner. Pieces of film were exposed to different doses in a dose range from 0.5 to 50 Gy using 6 MV photon beam. Functional forms for dose vs net optical density have been determined for each of the GafChromic film‐dosimetry systems used in this comparison. Two sources of uncertainties in dose measurements, governed by the experimental measurement and calibration curve fit procedure, have been compared for the densitometers used. Among the densitometers tested, it is found that for the HS film type the uncertainty caused by the experimental measurement varies from 1% to 3% while the calibration fit uncertainty ranges from 2% to 4% for doses above 5 Gy. Corresponding uncertainties for XR‐T film model are somewhat higher and range from 1% to 5% for experimental and from 2% to 7% for the fit uncertainty estimates. Notwithstanding the significant variations in sensitivity, the studied densitometers exhibit very similar precision for GafChromic film based dose measurements above 5 Gy.Keywords
Funding Information
- National Institutes of Health (P01-CA59827)
This publication has 26 references indexed in Scilit:
- The use of a reflective scanner to study radiochromic film responsePhysics in Medicine & Biology, 2002
- Evaluation of a fluorescent light densitometer for radiochromic film analysisRadiation Measurements, 2002
- Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributionsMedical Physics, 2000
- Quantitative optical densitometry with scanning-laser film digitizersMedical Physics, 1999
- Extrapolated surface dose measurements with radiochromic filmMedical Physics, 1999
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- Novel Radiochromic Films for Clinical DosimetryRadiation Protection Dosimetry, 1996
- Dosimetry systems for radiation processingRadiation Physics and Chemistry, 1995