Calibration and application of an intra-articular force transducer for the measurement of patellar tendon graft forces: an in situ evaluation.
- 1 August 1999
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 121 (4) , 393-398
- https://doi.org/10.1115/1.2798336
Abstract
The objective of this study was to evaluate two calibration methods for the "Arthroscopically Implantable Force Probe" (AIFP) that are potentially suitable for in vivo use: (1) a direct, experimentally based method performed by applying a tensile load directly to the graft after it is harvested but prior to implantation (the "pre-implantation" technique), and (2) an indirect method that utilizes cadaver-based analytical expressions to transform the AIFP output versus anterior shear load relationship, which may be established in vivo, to resultant graft load (the "post-implantation" technique). The AIFP outputs during anterior shear loading of the knee joint using these two calibration methods were compared directly to graft force measurements using a ligament cutting protocol and a 6 DOF load cell. The mean percent error (actual-measured)/(actual)* 100) associated with the pre-implantation calibration ranged between 85 and 175 percent, and was dependent on the knee flexion angle tested. The percent error associated with the post-implantation technique was evaluated in two load ranges: loads less than 40 N, and loads greater than 40 N. For graft force values greater than 40 N, the mean percent errors inherent to the post-implantation calibration method ranged between 20 and 29 percent, depending on the knee flexion angle tested. Below 40 N, these errors were substantially greater. Of the two calibration methods evaluated, the post-implantation approach provided a better estimate of the ACL graft force than the pre-implantation technique. However, the errors for the post-implantation approach were still high and suggested that caution should be employed when using implantable force probes for in vivo measurement of ACL graft forces.Keywords
This publication has 13 references indexed in Scilit:
- In-situ calibration of the implantable force transducerJournal of Biomechanics, 1996
- Evaluation of the implantable force transducer for chronic tendon-force recordingsJournal of Biomechanics, 1996
- In-vitro correlation between tension and length change in an anterior cruciate ligament substituteClinical Biomechanics, 1995
- Factors affecting sensitivity of a transducer for measuring anterior cruciate ligament forceJournal of Biomechanics, 1995
- Determination of a zero strain reference for the anteromedial band of the anterior cruciate ligamentJournal of Orthopaedic Research, 1994
- In vivo forces in the anterior cruciate ligament: Direct measurements during walking and trotting in a quadrupedJournal of Biomechanics, 1994
- In Vitro Evaluation of an Implantable Force Transducer (IFT) in a Patellar Tendon ModelJournal of Biomechanical Engineering, 1993
- Effect of tension and placement of a prosthetic anterior cruciate ligament on the anteroposterior laxity of the kneeJournal of Orthopaedic Research, 1992
- Anterior cruciate ligament replacements: A mechanical study of femoral attachment location, flexion angle at tensioning, and initial tensionJournal of Orthopaedic Research, 1990
- A Note on the Application and Evaluation of the Buckle Transducer for Knee Ligament Force MeasurementJournal of Biomechanical Engineering, 1982