Dosimetric characterization of a new design 103palladium brachytherapy source
- 11 November 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (11) , 2465-2470
- https://doi.org/10.1118/1.598765
Abstract
Low-energy gamma emitting isotopes encapsulated for permanent implant are in routine use in the treatment of prostate cancer. New source designs require full dosimetric analysis and calibration standardization before responsible clinical application. The results of one such experimental measurement and analysis are here reported for a new design of 103palladium source, model MED3633 (North American Scientific, Inc.). By AAMP Task Group #43 recommendations, the reference material for brachytherapy dosimetry is liquid water. The dose measurements were made using standard methods employing thermoluminescent dosimeters in a water equivalent plastic phantom. Precision machined bores in the phantom located dosimeters and source(s) in a reproducible fixed geometry providing for transverse-axis and angular dose profiles over a range of distances from 0.17 to 7 cm. The data were analyzed in terms of parameters recommended by AAPM TG43. The dose-rate constant, lambda, was evaluated with reference to a 60 cobalt standard, accounting for response variation with isotope energy spectrum. The radial dose function, g(r), the anisotropy function, F(r, theta), the anisotropy factor, phi,un(r), and the point-source approximation anisotropy constant, phi(un), were derived from one- and two-dimensional dose distribution data measured in the phantom, accounting for finite dosimeter volume and with attention to interchip effects. The results are compared to TG43 and other existing data for 103Pd sources. The new source is comparable to the model 200 103Pd source design, demonstrating equivalent radial dose function, g(r). The dose surrounding a MED3633 source may be slightly more isotropic than for the model 200 source. The air-kerma strength of the MED3633 source used in this study was provided by the manufacturer and is traceable to the NIST 1999 standard. The evaluated dose-rate constant, lambda, with NIST traceable strength calibration is lower than that of the model 200 source, with that manufacturer's strength calibration.Keywords
This publication has 8 references indexed in Scilit:
- Functional fitting of interstitial brachytherapy dosimetry data recommended by the AAPM Radiation Therapy Committee Task Group 43Medical Physics, 1999
- Anisotropy functions for and sourcesMedical Physics, 1998
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Diode dosimetry of 103 Pd model 200 seed in water phantomMedical Physics, 1994
- Comparison of radiation dosimetry in water and in solid phantom materials for I‐125 and Pd‐103 brachytherapy sources: EGS4 Monte Carlo studyMedical Physics, 1994
- Thermoluminescent dosimetry for 1 0 3 Pd seeds (model 200) in solid water phantomMedical Physics, 1991
- Dose parameters of 125I and 192Ir seed sourcesMedical Physics, 1989
- A comparison of solid phantoms with water for dosimetry of 125I brachytherapy sourcesMedical Physics, 1988