Ultrasound Attenuation Measurements of the Liver in Vivo Using a Commercial Sector Scanner
- 1 October 1984
- journal article
- Published by SAGE Publications in Ultrasonic Imaging
- Vol. 6 (4) , 396-407
- https://doi.org/10.1177/016173468400600402
Abstract
Attenuation measurements of various tissue minicking phantoms and three different groups of patients were obtained using a modified commercial sector scanner. Estimates of attenuation were made using the spectral shift method with mean frequencies at different depths of a region of interest being obtained by both zero crossing and fast Fourier transform techniques. The accuracy and precision of both techiques was compared in phantoms and it was found that the FFT technique yielded less day-to-day variation (SD=3 percent) than the zero crossing technique (5 percent). For larger regions of interest, the range of variation in both techniques was more similar. Day-to-day variation in livers of normal patients was much larger than that seen in phantoms (10 to 15 percent) suggesting that in vivo measurements may be less precise due to actual daily changes in patients' livers. Attenuation estimates of phantoms were high by approximately 0.16 dB/Mllz/cm compared to values obtained by transmission techniques. The attenuation values of livers in a group of 31 normal patients ranged from 0.214 dB/cm/MHz to 0.849 dB/cm/MHz with a mean of 0.627 ± 0.126 dB/cm/MHz for the zero crossing technique while the mean value using the FFT technique was 0.86 ± 0.168 dB/cm/MHz. A group of 26 Gauchers disease patients also showed wide variation with a mean attenuation value of 0.768 ± 0.21 dB/cm/MHz using the FFT technique. This was significantly different than that of the normal group (p<.05). Also, a group of 22 chronic B hepatitis patients was examined, having a mean attenuation value of 0.823 ± 0.21 dB/cm/MHz, not significantly different from those of the normal patients. Highly significant differences were found between the three groups when the power spectrum bandwidths of signals received were compared. These differences may be due to differences in the dependence of attenuation as a function of frequency between the groups and may represent a useful tissue characterization parameter.Keywords
This publication has 10 references indexed in Scilit:
- Estimation of the Slope of the Acoustic Attenuation CoefficientUltrasonic Imaging, 1984
- Attenuation of Ultrasound in Normal Liver and Diffuse Liver Disease in VivoUltrasonic Imaging, 1984
- Estimating acoustic attenuation from reflected ultrasound signals: Comparison of spectral-shift and spectral-difference approachesIEEE Transactions on Acoustics, Speech, and Signal Processing, 1984
- Improvements in the Spectral Difference Method for Measuring Ultrasonic AttenuationUltrasonic Imaging, 1983
- Spectral Characterization and Attenuation Measurements in UltrasoundUltrasonic Imaging, 1983
- Ultrasonic Attenuation Measurements of in Vivo Human MuscleUltrasonic Imaging, 1982
- Ultrasonography of alcoholic liver disease with histological correlation.Radiology, 1981
- Clinical Application of an Ultrasound Attenuation Coefficient Estimation Technique for Liver Pathology CharacterizationIEEE Transactions on Biomedical Engineering, 1980
- Accuracy of ultrasonography in diagnosis of hepatocellular diseaseAmerican Journal of Roentgenology, 1979
- Parametric Estimation of the Acoustic Attenuation Coefficient Slope for Soft TissuePublished by Institute of Electrical and Electronics Engineers (IEEE) ,1976