Dose reduction in double contrast barium enema by use of low fluoroscopic current
- 1 September 1996
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 69 (825) , 852-854
- https://doi.org/10.1259/0007-1285-69-825-852
Abstract
Under standard conditions double contrast barium enema (DCBE) was carried out on a selected patient cohort using either standard (1.5 mA) or low (0.5 mA) fluoroscopic current. A statistically significant (p≤0.01) reduction in total dose was achieved by use of low screening current. This reduction was of the order of 40%. Fluoroscopic image quality was analysed both objectively and subjectively, and was found not to be significantly degraded when the lower current was used. On blinded evaluation all examinations were found to be of diagnostic quality. The use of low fluoroscopic current results in a reduction in patient dose, without compromising fluoroscopic image quality or diagnostic standard of the examination. Low fluoroscopic current technique has been adopted as the standard in our department.Keywords
This publication has 5 references indexed in Scilit:
- Reduction of patient doses from barium meal and barium enema examinations through changes in equipment factorsThe British Journal of Radiology, 1994
- Estimation of effective dose from dose–area product measurements for barium meals and barium enemasThe British Journal of Radiology, 1994
- Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose-area productPhysics in Medicine & Biology, 1992
- Audit in action: significant reduction of double-contrast barium enema screening time with no loss of examination qualityThe British Journal of Radiology, 1992
- Patient dose reduction in diagnostic radiologyClinical Radiology, 1991