Mechanical characteristics of the stem‐cement interface
- 1 November 1991
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 9 (6) , 798-808
- https://doi.org/10.1002/jor.1100090605
Abstract
The mechanical characteristics of the interface between a metallic stem and the surrounding poly(methyl methacrylate) bone cement were determined from experimental tests and finite element analyses. Push-through-stem tests of straight and tapered titanium alloy stems, surrounded by cement columns, were performed and the resulting load-displacement behavior and strain distribution on the surface of the cement column were measured for loading, unloading, and reloading. Test geometries were modelled using nonlinear, axisymmetric, finite element analyses, which incorporated Coulomb friction elements at the titanium alloy-cement interface. Initial residual stresses, due to curing of the cement column, were modeled by thermal contraction of the cement. Good agreement was obtained between load-displacement curves and surface strains predicted from the nonlinear analysis and those obtained from experiments, when a coefficient of friction of 0.3 was assumed for the stem-cement interface. These results show that, in the absence of chemical adhesion, the load-displacement behavior of a stem-cement composite can be described completely in terms of the friction at the interface and the residual stresses normal to the interface.Keywords
This publication has 14 references indexed in Scilit:
- Cemented Total Hip Arthroplasty with Contemporary TechniquesClinical Orthopaedics and Related Research, 1988
- The effect of surface preparation on metal/bone cement interfacial strengthJournal of Biomedical Materials Research, 1987
- The load carrying and fatigue properties of the stem–cement interface with smooth and porous coated femoral componentsJournal of Biomedical Materials Research, 1985
- Metal/cement interface strength in cemented stem fixationJournal of Orthopaedic Research, 1984
- Thin film PMMA precoating for improved implant bone-cement fixationJournal of Biomedical Materials Research, 1982
- Intramedullary fixation of artificial hip joints with bone cement-precoated implants. I. Interfacial strengthsJournal of Biomedical Materials Research, 1982
- Transient and Residual Stresses and Displacements in Self-Curing Bone Cement—Part II: Thermoelastic Analysis of the Stem Fixation SystemJournal of Biomechanical Engineering, 1982
- Transient and Residual Stresses and Displacements in Self-Curing Bone Cement—Part I: Characterization of Relevant Volumetric Behavior of Bone CementJournal of Biomechanical Engineering, 1982
- The quasistatic and fatigue performance of the implant/bone-cement interfaceJournal of Biomedical Materials Research, 1981
- Factors affecting surgical alloy/bone cement interface adhesionJournal of Biomedical Materials Research, 1980