A Sample Purification Method for Rugged and High-Performance DNA Sequencing by Capillary Electrophoresis Using Replaceable Polymer Solutions. A. Development of the Cleanup Protocol
- 18 March 1998
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 70 (8) , 1516-1527
- https://doi.org/10.1021/ac971143f
Abstract
A method for the cleanup of Sanger DNA sequencing reaction products for capillary electrophoresis analysis with replaceable polymer solutions has been developed. A poly(ether sulfone) ultrafiltration membrane pretreated with linear polyacrylamide was first used to remove template DNA from the sequencing samples. Then, gel filtration in a spin column format (two columns per sample) was employed to decrease the concentration of salts below 10 μM in the sample solution. The method was very reproducible and increased the injected amount of the sequencing fragments 10−50-fold compared to traditional cleanup protocols. Using M13mp18 as template, the resulting cleaned-up single DNA sequencing fragments could routinely be separated to more than 1000 bases with a base-calling accuracy of at least 99% for 800 bases. The method is simple and universal and can be easily automated. In the following paper, a systematic study to determine quantitatively the effects of the sample solution components such as high-mobility ions (e.g., chloride and dideoxynucleotides) and template DNA on the injected amount and separation efficiency of the sequencing fragments is presented.Keywords
This publication has 16 references indexed in Scilit:
- Fully Automated DNA Reaction and Analysis in a Fluidic Capillary InstrumentAnalytical Chemistry, 1997
- Rapid DNA Sequencing of More Than 1000 Bases per Run by Capillary Electrophoresis Using Replaceable Linear Polyacrylamide SolutionsAnalytical Chemistry, 1996
- A Capillary Array Gel Electrophoresis System Using Multiple Laser Focusing for DNA SequencingAnalytical Chemistry, 1996
- DNA sequencing using a four‐color confocal fluorescence capillary array scannerElectrophoresis, 1996
- Multiple capillary DNA sequencer that uses fiber-optic illumination and detectionElectrophoresis, 1996
- Separation of DNA sequencing fragments using an automated capillary electrophoresis instrumentElectrophoresis, 1996
- Simultaneous monitoring of DNA fragments separated by electrophoresis in a multiplexed array of 100 capillariesAnalytical Chemistry, 1994
- Micellar electrokinetic capillary chromatography of nucleic acid constituents and dinucleoside monophosphate photoproductsJournal of High Resolution Chromatography, 1994
- Stability of capillary gels for automated sequencing of DNAElectrophoresis, 1992
- Analysis of DNA restriction fragments and polymerase chain reaction products towards detection of the AIDS (HIV-1) virus in bloodJournal of Chromatography A, 1991