Accuracy and reproducibility of the measurement of actively circulating blood volume with an integrated fiberoptic monitoring system
- 1 May 1995
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 23 (5) , 885-893
- https://doi.org/10.1097/00003246-199505000-00017
Abstract
Bedside monitoring of circulating blood volume has become possible with the introduction of an integrated fiberoptic monitoring system that calculates blood volume from the changes in blood concentration of indocyanine green dye 4 mins after injection. The aim of this investigation was to compare the blood volume estimate of the integrated fiberoptic monitoring system (group 1) with the standard methods of blood volume measurement using Evans blue (group 2), and indocyanine green measured photometrically (group 3).Prospective laboratory study.Animal laboratory of a University's institute for experimental surgery.Eleven anesthetized, paralyzed, and mechanically ventilated piglets.A central venous catheter was used for the injection of the indicator dyes (Evans blue and indocyanine green). A fiberoptic thermistor catheter was advanced into the thoracic aorta. The fiberoptic catheter detects indocyanine green by reflection densitometry for the estimation of blood volume of the integrated fiberoptic monitoring system. Samples for the determination of Evans blue and indocyanine green concentrations were drawn from an arterial catheter in the femoral artery over a period of 17 mins after injection.Measurements were performed during normovolemia, hypovolemia (blood withdrawal of < or = 30 mL/kg), and hypervolemia (retransfusion of the withdrawn blood plus an infusion of 10% hydroxyethyl starch [45 mL/kg]). Linear regression, correlation, and bias were calculated for the comparison of the blood volume estimates by the fiberoptic monitoring system (group 1) vs. the total blood volume estimates using Evans blue (group 2) and indocyanine green (group 3): group 1 = 0.82.group 2-26 mL; r2 = 82.71%; r = .91; n = 40; group 1-group 2 +/- 1 SD = -435 +/- 368 mL; group 1 = 0.79.group 3 + 50 mL; r2 = 74.81%; r = .87; n = 28; group 1-group 3 +/- 1 SD = -506 +/- 374 mL.The results demonstrate that the blood volume estimate of the fiberoptic monitoring system (group 1) correlates closely with the total blood volume measurement using Evans blue (group 2) and indocyanine green (group 3). Trapped indicator in the packed red cell column after centrifugation of the blood samples may account for an overestimation of group 2 and group 3 of approximately 10% to 14%, but there still remains a proportional difference of 10% between group 1 vs. group 2 and vs. group 3. This difference is due to the longer mixing times of group 3 (16 mins) and group 2 (17 mins), during which they are distributed in slowly exchanging blood pools. It seems that the blood volume estimate of the fiberoptic monitoring system (group 1) represents the actively circulating blood volume and may be useful for bedside monitoring.Keywords
This publication has 14 references indexed in Scilit:
- TEMPORAL BLOOD VOLUME, HEMODYNAMIC AND OXYGEN TRANSPORT PATTERNS IN ARDSCritical Care Medicine, 1994
- Determination of circulating blood volume by measurement of indocyanine green dye in hemolysate: A preliminary studyLife Sciences, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A method for measurement of total circulating blood volume using indocyanine green.The Tohoku Journal of Experimental Medicine, 1986
- Determination of blood volume using indocyanine green (Cardio-Green®) dyeLife Sciences, 1968
- Serial Cardiac Output and Blood Volume Studies Following Cardiac Valve ReplacementCirculation, 1966
- A CRITICAL COMPARISON OF THE T-1824 DYE AND IODINATED ALBUMIN METHODS FOR PLASMA VOLUME MEASUREMENT 1Journal of Clinical Investigation, 1953
- Determination of Circulating Red Cell Volume by Radioactive ChromiumScience, 1950
- A Method of Blood Volume DeterminationActa Physiologica Scandinavica, 1940
- CLINICAL STUDIES OF THE BLOOD VOLUME. I. CLINICAL APPLICATION OF A METHOD EMPLOYING THE AZO DYE “EVANS BLUE” AND THE SPECTROPHOTOMETERJournal of Clinical Investigation, 1937