Clinical Evaluation of the Effects of Signal Integrity and Saturation on Data Availability and Accuracy of Masimo SET?? and Nellcor N-395 Oximeters in Children
- 1 March 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 98 (3) , 617-622
- https://doi.org/10.1213/01.ane.0000101601.43781.15
Abstract
Pulse oximetry manufacturers have introduced technologies that claim improved detection of hypoxemic events. Because improvements in signal processing and data rejection algorithms may differentially affect data reporting, we compared the data reporting and signal heuristic performance and agreement among the Nellcor N-395, Masimo SET, and GE Solar 8000 oximeters under a spectrum of conditions of signal integrity and arterial oxygen saturations. A blinded side-by-side comparison of technologies was performed in 27 patients, and data were analyzed for time of data availability, measures of agreement and signal heuristics, and warnings stratified by signal integrity and SpO(2). The Solar 8000 had less total data dropout than either of the new technologies. Masimo's LoSIQ (signal quality) heuristic rejected less data than Nellcor's MOT/PS (motion/pulse search) flag. When no signal heuristic was displayed, there was little difference in precision and bias between the two newer technologies; however, agreement between devices deteriorated in the presence of SIQ, MOT, or hypoxemia. Both newer devices flagged questionable data, but their use of different rejection algorithms resulted in different probabilities of presenting data. Therefore, with poor SIQ or during hypoxemia, the Nellcor N-395 and Masimo oximeters are not clinically equivalent to each other or to the older Solar 8000 oximeter. We compared new pulse oximeters from Nellcor and Masimo and found that, with good signal conditions, both new devices performed similarly to older technology. Overall, Masimo reported less data as questionable than Nellcor. With poor signal conditions or during hypoxemia, the new devices are not clinically equivalent to each other or to the older technology.Keywords
This publication has 19 references indexed in Scilit:
- “Motion-Resistant” Pulse Oximetry: A Comparison of New and Old ModelsAnesthesia & Analgesia, 2002
- Reliability of Conventional and New Pulse Oximetry in Neonatal PatientsJournal of Perinatology, 2002
- False Alarms and Sensitivity of Conventional Pulse Oximetry Versus the Masimo SET™ Technology in the Pediatric Postanesthesia Care UnitAnesthesia & Analgesia, 2000
- Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: Comparison between a conventional and two new generation oximetersCritical Care Medicine, 2000
- Masimo Signal Extraction Pulse OximetryJournal of Clinical Monitoring and Computing, 2000
- Effects of motion, ambient light, and hypoperfusion on pulse oximeter functionJournal of Clinical Anesthesia, 1997
- Postanesthesia monitoring revisited: Frequency of true and false alarms from different monitoring devicesJournal of Clinical Anesthesia, 1994
- Crying wolfCritical Care Medicine, 1994
- Randomized Evaluation of Pulse Oximetry in 20,802 Patients; IAnesthesiology, 1993
- A prospective study of intraoperative pulse oximetry failureJournal of Clinical Monitoring and Computing, 1991