TEXT heavy ion beam probe system
- 1 August 1986
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 57 (8) , 1825-1827
- https://doi.org/10.1063/1.1139141
Abstract
A good signal-to-noise ratio has been obtained during initial operation of the 500-keV TEXT heavy ion beam probe. A number of problems were identified during the start up phase and they have been or are being corrected. Present sensitivity for detecting changes in the space potential is 50 V, but it is expected that this can be reduced to 5 V when additional calibration and alignment procedures are completed.Keywords
This publication has 5 references indexed in Scilit:
- Space Potential Distribution in the ISX-B TokamakPhysical Review Letters, 1986
- Comparison of the Radial Potential Profile Measured in a Tokamak to Predictions of Stochastic Magnetic Field TheoryPhysical Review Letters, 1985
- Effects of intense ultraviolet radiation on electrostatic energy analyzersReview of Scientific Instruments, 1985
- Minimum Image Size in a Parallel Plate Electrostatic SpectrographReview of Scientific Instruments, 1970
- A Parallel Plate Electrostatic SpectrographReview of Scientific Instruments, 1970