Virtual electrode polarization in the far field: implications for external defibrillation
- 1 September 2000
- journal article
- research article
- Published by American Physiological Society in American Journal of Physiology-Heart and Circulatory Physiology
- Vol. 279 (3) , H1055-H1070
- https://doi.org/10.1152/ajpheart.2000.279.3.h1055
Abstract
We recently suggested that failure of implantable defibrillation therapy may be explained by the virtual electrode-induced phase singularity mechanism. The goal of this study was to identify possible mechanisms of vulnerability and defibrillation by externally applied shocks in vitro. We used bidomain simulations of realistic rabbit heart fibrous geometry to predict the passive polarization throughout the heart induced by external shocks. We also used optical mapping to assess anterior epicardium electrical activity during shocks in Langendorff-perfused rabbit hearts (n= 7). Monophasic shocks of either polarity (10–260 V, 8 ms, 150 μF) were applied during the T wave from a pair of mesh electrodes. Postshock epicardial virtual electrode polarization was observed after all 162 applied shocks, with positive polarization facing the cathode and negative polarization facing the anode, as predicted by the bidomain simulations. During arrhythmogenesis, a new wave front was induced at the boundary between the two regions near the apex but not at the base. It spread across the negatively polarized area toward the base of the heart and reentered on the other side while simultaneously spreading into the depth of the wall. Thus a scroll wave with a ribbon-shaped filament was formed during external shock-induced arrhythmia. Fluorescent imaging and passive bidomain simulations demonstrated that virtual electrode polarization-induced scroll waves underlie mechanisms of shock-induced vulnerability and failure of external defibrillation.Keywords
This publication has 31 references indexed in Scilit:
- Virtual Electrode–Induced ReexcitationCirculation Research, 1999
- Evidence of three-dimensional scroll waves with ribbon-shaped filament as a mechanism of ventricular tachycardia in the isolated rabbit heart.Journal of Cardiovascular Electrophysiology, 1999
- High-resolution, three-dimensional fluorescent imaging reveals multilayer conduction pattern in the atrioventricular node.Circulation, 1998
- Virtual Electrode–Induced Phase SingularityCirculation Research, 1998
- A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular ArrhythmiasNew England Journal of Medicine, 1997
- Transmembrane Voltage Changes Produced by Real and Virtual Electrodes During Monophasic Defibrillation Shock Delivered by an Implantable ElectrodeJournal of Cardiovascular Electrophysiology, 1997
- High‐Resolution Fluorescent Imaging Does Not Reveal a Distinct Atrioventricular Nodal Anterior Input Channel (Fast Pathway) in the Rabbit Heart During Sinus RhythmJournal of Cardiovascular Electrophysiology, 1997
- Optical mapping of repolarization and refractoriness from intact hearts.Circulation, 1994
- Potential Distributions Generated By Point Stimulation in a Myocardial Volume:Journal of Cardiovascular Electrophysiology, 1993