Surface Damage on Open Box Posterior-Stabilized Polyethylene Tibial Inserts
- 1 November 2003
- journal article
- section i-symposium
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 416 (416) , 135-144
- https://doi.org/10.1097/01.blo.0000092998.90435.98
Abstract
Twenty retrieved Scorpio® posterior-stabilized implants were available for analysis. The mean implantation time was 22 months (range, 2 days–42 months). Favorable types and amounts of surface damage were seen on the tibiofemoral and backsides of these modular PE liners that had been packaged in an inert environment and then sterilized by gamma irradiation. Delamination represented only 0.1% of the total surface damage. Off-axis loading (varus malalignment) was associated with tibial component loosening but there was no evidence of peripheral damage of PE caused by edge loading. With this open-box design, hyperextension marks on the anterior aspect of the posterior-stabilized post from femoral component impingement occurred in 11 of 20 cases and was related to sagittal component positioning: excess tibial slope or increased tibial slope combined with a flexed femoral component. Unique coarse abrasions occurred in 16 of 20 cases and were the result of cement extrusion into the open box, especially with varus malalignment. These observations provide guidance for optimizing the surgical technique and the design of posterior-stabilized total knee components.Keywords
This publication has 10 references indexed in Scilit:
- SURFACE DAMAGE IN MACHINED RAM-EXTRUDED AND NET-SHAPE MOLDED RETRIEVED POLYETHYLENE TIBIAL INSERTS OF TOTAL KNEE REPLACEMENTSJournal of Bone and Joint Surgery, 2002
- The effect of total knee arthroplasty design on extensor mechanism functionThe Journal of Arthroplasty, 2002
- In Vivo Deterioration of Tibial Baseplate Locking Mechanisms in Contemporary Modular Total Knee ComponentsJournal of Bone and Joint Surgery, 2001
- Tibial Post Wear in Posterior Stabilized Total Knee ArthroplastyJournal of Bone and Joint Surgery, 2001
- Loosening and Osteolysis with the Press-Fit Condylar Posterior-Cruciate-Substituting Total Knee ReplacementJournal of Bone and Joint Surgery, 2001
- The conflicting requirements of laxity and conformity in total knee replacementJournal of Biomechanics, 1999
- Current Concepts Review - Wear in Total Hip and Knee Replacements*Journal of Bone and Joint Surgery, 1999
- Modular Tibial Insert MicromotionPublished by Wolters Kluwer Health ,1998
- Impact of gamma sterilization on clinical performance of polyethylene in the kneeThe Journal of Arthroplasty, 1996
- The Knee Society Total Knee Arthroplasty Roentgenographic Evaluation and Scoring SystemClinical Orthopaedics and Related Research, 1989