Assessment of Titanium Alloy on Polyethylene Bearing Surfaces in Retrieved Uncemented Total Hip Replacements
- 1 June 1994
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 208 (2) , 91-95
- https://doi.org/10.1243/pime_proc_1994_208_270_02
Abstract
Changes to the bearing surfaces of eighteen uncemented total hip replacements retrieved at revision surgery were assessed by three-dimensional binocular microscopy, Rank Taylor Hobson talysurf measurements, scanning electron microscopy, and X-ray dispersive analysis. Abrasions on the non-articular surface of the polyethylene cups were present. Bone particles were found in tracks in the bearing surfaces of both the titanium femoral heads and the polyethylene cups and were responsible for wear of these surfaces. Although the wear of the femoral heads appeared substantial to naked eye examination, the surface finish of these surfaces remained within the British ISO standards for titanium alloy when assessed by the methods used above. This study concluded that direct contact between polyethylene and bone should be avoided in total hip arthroplasty and that ‘third body’ wear from bone particles occurred in these uncemented prostheses. Both components of this type of implant should be replaced at revision surgery and titanium should be avoided as a bearing surface in hip arthroplasty. Present methods of assessing the surface finish of titanium should be re-evaluated and more reliable ones considered.Keywords
This publication has 22 references indexed in Scilit:
- Ceramic bearing surfaces in total artificial joints: Resistance to third body wear damage from bone cement particlesJournal of Medical Engineering & Technology, 1991
- Cementless Total Joint Arthroplasty Prostheses With Titanium-Alloy Articular SurfacesPublished by Wolters Kluwer Health ,1990
- The Biomechanical Problems of Polyethylene as a Bearing SurfaceClinical Orthopaedics and Related Research, 1990
- Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year resultsThe Journal of Bone and Joint Surgery. British volume, 1990
- The Influence of Bone and Bone Cement Debris on Counterface Roughness in Sliding Wear Tests of Ultra-High Molecular Weight Polyethylene on Stainless SteelProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990
- Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysis.Journal of Bone and Joint Surgery, 1988
- The role of counterface imperfections in the wear of polyethyleneWear, 1987
- Uncemented Acetabular Cups in Dysplastic and Protrusio AcetabuliPublished by Wolters Kluwer Health ,1986
- Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: III: The measurement of internal volume changes in explanted Charnley sockets after 2–16 years in vivo and the determination of wear factorsWear, 1985
- The Pathology of Failed Total Joint ArthroplastyPublished by Wolters Kluwer Health ,1982