Multi‐gradient echo with susceptibility inhomogeneity compensation (MGESIC): Demonstration of fMRI in the olfactory cortex at 3.0 T
- 1 March 1997
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 37 (3) , 331-335
- https://doi.org/10.1002/mrm.1910370304
Abstract
Short image acquisition times and sensitivity to magnetic susceptibility favor the use of gradient echo imaging methods in functional MRI (fMRI). However, magnetic susceptibility effects attributed to air-tissue interfaces also lead to severe signal loss in images of the large inferior frontal and lateral temporal cortices of the human brain, which renders these regions inaccessible to fMRI. The signal loss is caused by the local field gradients in the slice selection direction. A multigradient echo with magnetic susceptibility inhomogeneity compensation method (MGESIC) is proposed to overcome this problem. The MGESIC method effectively corrects the susceptibility artifacts and maintains the advantages of gradient echo methods to both BOLD sensitivity and fast image acquisition. The effectiveness of the MGESIC method is demonstrated by fMRI experimental results within the olfactory cortex.Keywords
This publication has 17 references indexed in Scilit:
- Functional 2D and 3D magnetic resonance imaging of motor cortex stimulation at high spatial resolution using standard 1.5 T imagerMagnetic Resonance Imaging, 1994
- Assessment of relative brain iron concentrations using T2‐weighted and T2*‐weighted MRI at 3 TeslaMagnetic Resonance in Medicine, 1994
- Spatial patterns of functional activation of the cerebellum investigated using high field (4 T) MRINMR in Biomedicine, 1994
- 2D and 3D high resolution gradient echo functional imaging of the brain: Venous contributions to signal in motor cortex studiesNMR in Biomedicine, 1994
- Functional brain imaging at 1.5 T using conventional gradient echo MR imaging techniquesMagnetic Resonance Imaging, 1993
- Functional mapping of the human visual cortex at 4 and 1.5 tesla using deoxygenation contrast EPIMagnetic Resonance in Medicine, 1993
- Functional MRI of human brain activation at high spatial resolutionMagnetic Resonance in Medicine, 1993
- Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging.Proceedings of the National Academy of Sciences, 1992
- Reduction of susceptibility artifact in gradient‐echo imagingMagnetic Resonance in Medicine, 1992
- Direct FLASH MR imaging of magnetic field inhomogeneities by gradient compensationMagnetic Resonance in Medicine, 1988