Counting platelets in platelet concentrates on hematology analyzers: a multicenter comparative study
- 23 December 2008
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 49 (1) , 81-90
- https://doi.org/10.1111/j.1537-2995.2008.01930.x
Abstract
Hematology analyzers are designed to count whole blood samples, but are also used by blood centers to perform quality control on blood components. In platelet (PLT) concentrates, the number of PLTs is approximately fivefold higher and red blood cells are absent, causing variable PLT counting results. It was our aim to compare currently used hematology analyzers for counting PLTs in PLT concentrates using fixed human PLTs. PLT samples were fixed, diluted into seven concentration levels (plus one blank), aliquoted, and shipped to 68 centers. Evaluable data were obtained for 89 hematology analyzers. All samples were counted six times, and results were reported to the coordinating center. The overall group mean was calculated, and the percentage deviation from this mean was calculated for each analyzer. At PLT levels relevant for blood centers, 750 x 10(9) to 2000 x 10(9) per L, analyzers gave results that were between 35 percent lower and 16 percent higher than the overall group mean. Within a group of analyzers, results were comparable with coefficient of variations usually below 10 percent, indicating that the observed differences were caused by instrument characteristics. A smaller study with fresh, unfixed PLT samples showed that analyzers behaved similarly for fixed and fresh PLTs. With a wide array of currently used hematology analyzers, a marked difference was determined for the PLT counts of fixed human-based identical samples provided to 68 laboratories by a centralized facility. A gold standard method is needed to allow for more valid interlaboratory comparisons between hematology analyzers.Keywords
This publication has 11 references indexed in Scilit:
- Platelet counting in platelet concentrates with various automated hematology analyzersTransfusion, 2007
- The Influence of Various Hematology Analyzers on Component Platelet CountsTransfusion Medicine Reviews, 2005
- Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusionBritish Journal of Haematology, 2005
- The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet componentsVox Sanguinis, 2004
- Standardisation of platelet counting accuracy in blood banks by reference to an automated immunoplatelet procedure: comparative evaluation of Cell-Dyn CD4000 impedance and optical platelet countsTransfusion and Apheresis Science, 2001
- Platelet Counting by the RBC/Platelet Ratio MethodAmerican Journal of Clinical Pathology, 2001
- Immunoplatelet counting: a proposed new reference procedureBritish Journal of Haematology, 2000
- An evaluation of the CELL-DYN 1700R haematology analyser: automated cell counting and three-part leucocyte differentiationInternational Journal of Laboratory Hematology, 1996
- Reliability of automated platelet counts: Comparison with manual method and utility for prediction of clinical bleedingAmerican Journal of Hematology, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986